Nidek Medical Mark 5 Nuvo 10 User Manual Download Page 3

                                   November 2013

2010-2324CE Rev A

3

3

4

5

 

I.1. Front panel (Fig. I.1)

   1  -

 (

I/0

) On/Off Power Switch 

   2 

-  Humidifier (space reserved)

   3  -

  Oxygen product outlet 

   4 - 

  Flow adjustment knob (l/min) 

   5 

-  Circuit Breaker

   6 

-  OCSI (Oxygen Concentration    

 

 

   

Status Indicator) Indicator Lights (if equipped)

I.2. Rear panel (Fig. I.2)

 

 7  - 

Cabinet Air Filter

 8  -

 

Manufacturer's Technical Label

 9  - 

Power Cord 

8

3

2

3

5

1

  5

4

6

3

    

       

     

 

4

    

6

    

 

 

3

    

  5

7

7

9

2

1

II

.  STARTING UP / INSTALLATION

II.1. Use in direct oxygen therapy

a. Ensure that the switch (1) is in the 

0

 

(OFF) position.

b. If used with a humidifier:

Unscrew the flask and fill it with water up  

to the line (see humidifier instructions). Then 

screw the lid on the humidifier flask until there 

are no leaks.

c. Connect the oxygen tube to the humidifier outlet nozzle 

or  to  the  concentrator  outlet  if  a  humidifier  has  not  been 

prescribed. The tube between the cannula and the

 

Mark 5 

Nuvo MB

 

should be limited to 20 meters (60 feet) long

in order to ensure that the oxygen flow rate remains within 

specification values. 

d. Ensure that all of the parts are connected correctly so as 

to avoid leaks.

e. Plug the power cable into a power outlet 

of the correct   voltage  and  frequency  as 

defined on the  manufacturer's  technical 

label (8).           

                   

            

  

f. Press the power switch 

( I / 0 )

 

 to the ON position

 ( I )

.  

The green indicator will light when the oxygen concentration 

exceeds the set point (OCSI) or normal operating pressures. 

Note:

 After turning the unit off, the user must wait 3-5 

minutes before turning it back on. System pressure must 

dissipate before the unit will properly restart. 

g. Turn the flow adjustment knob (4) to the prescribed value. 

This knob may have already been locked in the medically 

prescribed position. In this case, do not force it. Only the 

technician  or  medical  personnel  are  authorized  to  release 

it.   

Note

: View the flowmeter in the horizontal plane for 

accurate settings.

h.  Check  that  the  oxygen  flows  out  of  the  administration 

device (nasal cannulas or other) by placing the orifice(s) on 

the surface of a glass of water. The flow should disturb the 

surface of the water.

i. Adjust the nasal cannula to suit your face.

Note

: the required oxygen concentration is normally obtained 

within five minutes after the unit is started.

 

At the end of the treatment, press the 

I/0

 Switch  to place it 

in the 

(OFF) position to stop the device. The oxygen enri

-

ched air flow continues for approximately one minute after 

the device is stopped.

For the equipment supplier or medical staff:

Summary of Contents for Mark 5 Nuvo 10

Page 1: ...3 III CLEANING MAINTENANCE 4 III 1 Cleaning 4 III 2 Everyday disinfection 4 III 3 Maintenance 4 IV USEFUL INFORMATION 4 IV 1 Accessories and spare parts 4 IV 2 Materials in direct indirect contact wi...

Page 2: ...r installer or distributor are not considered to be responsible themselves for the consequences on the safety reliability and characteristics of a device unless Theassembly fitting extensions adjustme...

Page 3: ...of the correct voltage and frequency as defined on the manufacturer s technical label 8 f Press the power switch I 0 to the ON position I The green indicator will light when the oxygen concentration e...

Page 4: ...ons The cabinet air filter should be washed or replaced The entire oxygen administration circuit oxygen therapy nasal cannulas etc must be changed III 3 Maintenance No special maintenance needs to be...

Page 5: ...angle transport with humidifier fitted 70o Weight 24 kg 54 lbs Noise level within ISO 8359 guidelines Flow values Continuously Adjustable Flowmeter 1 to 5 2 to 8 or 2 to 10 depending on model IV 5 3...

Page 6: ...the maximum recommended flow does not exceed 10 of the indicated value when a back pressure of 6 9 kPa 1 psig is applied to the output of the device The maximum outlet pressure is 7psig for 5 l min 15...

Page 7: ...vice provider for any updates to recommended schedules The work must be carried out by suitably trained technicians certified by the manufacturer Use original spare parts only see Pg 8 Uponrequest the...

Page 8: ...in the I ON position but the green indicator is not lighted Faulty indicator Contact your equipment supplier The I 0 ON OFF button is in the I ON position but there is no flow The audible alarm sound...

Page 9: ...November 2013 2010 2324CE Rev A 9 V EMC Electromagnetic Statements...

Page 10: ...010 2324CE Rev A 10 November 2013 NOTES...

Page 11: ...November 2013 2010 2324CE Rev A 11 NOTES...

Page 12: ...alley East Industrial Drive Birmingham Alabama 35217 U S A Tel 205 856 7200 Fax 205 856 0533 EU Representative mdi Europa GmbH Langenhagener Str 71 30855 Hannover Langenhagen Germany Tel 49 511 39 08...

Reviews: