Nidek Medical Mark 5 Nuvo 10 User Manual Download Page 1

                                 November 2013

2010-2324CE Rev A

1

USER'S   GUIDE

Mark 5 Nuvo

 

with Modular Bed

  

NUVO, NUVO 8 & NUVO 10  

OXYGEN  CONCENTRATOR 

Federal Law (US) restricts this device to sale 

by, or on the order of, a licensed  physician.  

This oxygen concentrator should be used 

only under the supervision of a licensed 

physician. 

 

        

 

0413

:  

Complies with the 93/42/ EEC directive 

certified by the approved organization no 0413.

Danger:  Do not  smoke when using oxygen or  when 

near this device.

GLOSSARY OF SYMBOLS ..........................................1
GENERAL SAFETY GUIDELINES ............................2      
I.     DESCRIPTION .......................................................2

I.1.  

F

ront panel  (Fig. I. 1)

 ........................................... 3

I.2.  Rear panel (Fig. I. 2) ..................................................3

II.   STARTING-UP / INSTALLATION .......................3

II.1. 

Use in direct oxygen therapy .....................................3

III.  CLEANING-MAINTENANCE .............................4

III.1.Cleaning

 ....................................................................4

III.2.Everyday disinfection ....................................................4

III.3.Maintenance ..............................................................4

IV.USEFUL INFORMATION  .........................................4

IV.1.  Accessories and spare parts

 ....................................... 4

IV.2.  Materials in direct/indirect contact with patient........ 4

IV.3.  Operating principles ..................................................5

IV.4.  Alarms - Safety devices

 ............................................ 5

IV.5.

 

Oxygen Monitor (OCSI)

 

function

 

  ...........................

5

IV.6.  Technical characteristics ........................................... 5

IV.7. Standards ................................................................... 6

IV.8.  Method for disposing of waste .................................. 6

IV.9.  Method for disposing of device................................. 6

IV.10.Troubleshooting

......................................................... 7

V. EMC, Electromagnetic statements

 ...........................

8-9

CONTENTS

  

: ON (power switched on)  

: Off (power switched off) 

  

   

: Type B device

 : Class II protection
  

  

 

: Do not expose to open flames

 

: Do not use oil or grease

: Technical information

: Consult the accompanying documents

: Keep in the vertical position

: Fragile - handle with care

: Oxygen concentration warning light

GLOSSARY OF SYMBOLS

[Original language is English]

Summary of Contents for Mark 5 Nuvo 10

Page 1: ...3 III CLEANING MAINTENANCE 4 III 1 Cleaning 4 III 2 Everyday disinfection 4 III 3 Maintenance 4 IV USEFUL INFORMATION 4 IV 1 Accessories and spare parts 4 IV 2 Materials in direct indirect contact wi...

Page 2: ...r installer or distributor are not considered to be responsible themselves for the consequences on the safety reliability and characteristics of a device unless Theassembly fitting extensions adjustme...

Page 3: ...of the correct voltage and frequency as defined on the manufacturer s technical label 8 f Press the power switch I 0 to the ON position I The green indicator will light when the oxygen concentration e...

Page 4: ...ons The cabinet air filter should be washed or replaced The entire oxygen administration circuit oxygen therapy nasal cannulas etc must be changed III 3 Maintenance No special maintenance needs to be...

Page 5: ...angle transport with humidifier fitted 70o Weight 24 kg 54 lbs Noise level within ISO 8359 guidelines Flow values Continuously Adjustable Flowmeter 1 to 5 2 to 8 or 2 to 10 depending on model IV 5 3...

Page 6: ...the maximum recommended flow does not exceed 10 of the indicated value when a back pressure of 6 9 kPa 1 psig is applied to the output of the device The maximum outlet pressure is 7psig for 5 l min 15...

Page 7: ...vice provider for any updates to recommended schedules The work must be carried out by suitably trained technicians certified by the manufacturer Use original spare parts only see Pg 8 Uponrequest the...

Page 8: ...in the I ON position but the green indicator is not lighted Faulty indicator Contact your equipment supplier The I 0 ON OFF button is in the I ON position but there is no flow The audible alarm sound...

Page 9: ...November 2013 2010 2324CE Rev A 9 V EMC Electromagnetic Statements...

Page 10: ...010 2324CE Rev A 10 November 2013 NOTES...

Page 11: ...November 2013 2010 2324CE Rev A 11 NOTES...

Page 12: ...alley East Industrial Drive Birmingham Alabama 35217 U S A Tel 205 856 7200 Fax 205 856 0533 EU Representative mdi Europa GmbH Langenhagener Str 71 30855 Hannover Langenhagen Germany Tel 49 511 39 08...

Reviews: