TNP Digital Torque Meter

Operation Manual

Do not attempt to repair or alter this instrument.  Warranty will be voided 

and damage to the unit may result.

Use and store within the stated temperature and humidity ranges, or 

damage and failure may result.

Ensure during testing unit is properly used on a stable, level surface. 

Do not test products that are filled with liquid as the TNP torque tester is 
not protected from liquid spills.

Only use electrical charger supplied with unit.

SPECIFICATIONS
Measuring Range: TNP-0.5:

 0-50.00 Ncm, 0-500.0 mNm, 

0-5099 gcm, 0-4.425 lbin; 

TNP-2:

 0-2.000 Nm, 0-200.0 Ncm, 

0-20.39 kgcm, 0-17.70 lbin; 

TNP-5:

 0-5.00 Nm, 0-500.0 Ncm, 

0-50.99 kgcm, 0-44.25 lbin; 

TNP-10:

 0-10.00 Nm, 0-1000 Ncm, 

0-102.0 kgcm, 0-88.5 lbin

Test Sample Diameter Range: 

0.39“ - 7.48” (10 - 190 mm)

Accuracy: 

±0.5% / F.S.

Display: 

Main display: 4-digit LCD display, Character height 0.5” 

(12 mm)

Comparator Display: 

HI - GO - LO (green and red LED indica-

tor)

Display Update Time: 

Select from 0.125 second (8 times/sec-

ond), 0.25 second (4 times/second), 0.5 second (2 times/second), 
1 second (1 time/second). Peak display: 0.125 second

Measuring Mode: 

Opening Mode: Show max value when open-

ing (peak display) Displays max counterclockwise torque; Closing 
Mode: Show max value when closing (peak display) Displays max 
clockwise torque; Average Mode: Real time display (average value 
display) Displays torque in real time

Overload: 

150% of full scale

Sampling Rate: 

1000 Hz

Memory: 

1000 data (max)

Data Output: 

USB1.1

Power: 

Built in nickel hydride battery (NiMH) or Auto-ranging AC 

adapter (AC100 - 240V)

Battery Information: 

NiMH: 8 hours use after full charge. 16 

hour max. recharge time.

Operating Environment: 

32 - 104°F (0 - 40°C)

Dimensions: 

12.60” x 8.94” x 4.96” (320 x 227x 126 mm)

Product Weight: 

TNP-0.5, -2, -5 = Approx. 17.6 lb (8 kg);   TNP-

10 = 27.5 lb (12.5 kg)

Package Weight: 

TNP-0.5, -2, -5 = Approx. 19 lb (8.6 kg);   TNP-

10 = 30.7 lb (13.9 kg)

Software: 

DigiTorque TNP Free Download from www.shimpoinst.

com

Included Accessories: 

USB cable, 120/240 VAC AC adapter, 

30mm chuck pin (4pcs)

Certifications: 

CE

Warranty:

 2 year

The Series TNP Digital Torque Meter is a precision instrument for 
measuring OPEN and CLOSE torque values for various applica-
tions.

 

The TNP Torque Meter is used in medical, pharmaceutical, 

industrial quality control, R&D and countless other torque appli-
cations.

As a cap torque analyzer, the TNP repeatedly determines bottle 
closure and breakaway torque. Its wide memory capacity is ca-
pable of storing 1,000 data points, including Average, Maximum 
and Minimum values. A HI-GO-LO LED comparator feature can be 
utilized for large batch testing to ensure overall product quality. 
The TNP is a portable device powered with internal rechargeable 
batteries or through provided universal AC adapter/charger.  

Designed with a small footprint, it can easily be transferred be-
tween the shop floor and laboratory taking up little space when 
operating. The TNP Torque Meter comes standard with USB con-
nection and cable for easy PC/Software interface. Combined with 
the free “Digitorq” Software, statistical testing, graphing and anal-
ysis can be performed.

Each package includes the following:
• Operations Manual
• Software Installation Disk (Digitorq Software)
• Warranty Card
• USB A to USB B communication cable
• Universal AC adapter (100-240 VAC)
• Set of 30 mm chuck pins (4 pieces)
• TNP unit

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for TNP Series

Page 1: ...charge 16 hour max recharge time Operating Environment 32 104 F 0 40 C Dimensions 12 60 x 8 94 x 4 96 320 x 227x 126 mm Product Weight TNP 0 5 2 5 Approx 17 6 lb 8 kg TNP 10 27 5 lb 12 5 kg Package W...

Page 2: ...in parameter settings see next section Zero RST key Tare or zero function for resetting and initializing values while in Average or peak mode Secondary function in parameter settings see next section...

Page 3: ...before the unit turns off Main display Set up Values shows the measured value in four digits including decimal places While under function mode this displays the options for each setting Direction In...

Page 4: ...que meter is in Function Mode Operation key Operation How to operate Zero RST Power Function mode With Power off press and hold the Zero RST key then power ON the torque meter Continue holding the ZER...

Page 5: ...entation F04 0000 CCW Open 0000 CCW Open 0000 Upper comparator value Hi limit HI 0000 9999 with decimal point Setting the values to zero disables this function 0000 Lower comparator value Lo limit LO...

Page 6: ...ease the Power button but continue to hold Zero RST key until display F01 Function Mode Measurement unit Power Zero RST Unit N m Unit N cm Ncm Nm Unit Kg cm Kgc Unit lb in lbi TNP 2 5 10 Mode Mode Mod...

Page 7: ...n Mode Auto power OFF Disable auto power OFF Mode Mode Zero RST Pressing Zero RST it moves to the next function setting f04 Open f04 Function Mode Sign Orientation Open Mode Mode Zero RST Pressing Zer...

Page 8: ...all Recall Mem Mem Recall Zero RST Pressing Zero RST it moves to the next function setting Function Mode Lower comparator value Lo Limit Key operation is the same in entering the Hi limit value LO Pre...

Page 9: ...mode required for testing Please check the settings below on how to select the right mode for testing Pressing the Mode key toggles through the mode available from the TNP LED indicators are provided...

Page 10: ...O HI LO HI Red LED indicator will be lit on the Comparator LED indicators Utilizing this feature makes it an ideal tool for quality assurance checks Example HI is set to 100 and LO is set to 50 Based...

Page 11: ...F01 Press the Zero RST key until the main display shows HI limit Using the Recall key we can increment the highlighted digit from 0 9 Press the Mem key to move from left to right or to the next digit...

Page 12: ...ta can be stored in the TNP meter by pressing the Mem key This feature is only available in OPEN and CLOSE measuring modes NOTE In MEAN or Average mode the Mem key has no function When the number of d...

Page 13: ...es after pressing Recall To access the stored measured value use the Recall key to scroll thru the values This will be after the MIN value is displayed The display will flash two sets of numbers the f...

Page 14: ...X Maximum value CNT Alternating display 1sec interval Shows last memory data Returns to measuring mode Nm Nm Nm Alternating display 1sec interval Nm Nm Test number displayed Last memory number data di...

Page 15: ...e memory to be deleted jump out of sequence How does it work First select from Open or Close Modes Mean or average mode data cannot be stored Press the Recall key to access memory the first display yo...

Page 16: ...er off the TNP torque meter Press the Mem key and continue to hold this key while powering the torque tester The TNP will initialize itself You will see the model capacity displayed on the front panel...

Page 17: ...DIMENSIONS 17 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Reviews: