background image

Thank you for purchasing NIDEC COPAL ELECTRONICS 
CORP. product.
In order to use the product correctly and most appropriately, 
please  completely  read  this  manual  before  use  and  keep  it 
for future reference.

1

This caution mark indicates a chance of physical damages to the user and/or the product 

if it is used improperly and/or incorrectly. The damage can be substantial. 

HANDY MANOMETER

PG-100B

Series

Important Information and Warnings

Operating Precautions

(1)The applicable media are gases/liquids compatible with SUS316L.

Materials of the pressure port are SUS316L while those of the attachment are SUS316 and SUS 304. So please be careful about compatibility with 

applicable media. 

Use of corrosive fluids may cause injury and/or malfunction of products.

In addition, the product is not explosive-proof. Never use flammable fluid as media.

(2)Maintain the measuring pressure within the pressure range as stated in specifications.

Pressures exceeding the maximum pressure rating may cause product failure.

(3)Be careful not to let any liquid inside the main product.

Liquid leak on the pressure port may cause product failure. So please be careful to remove fluids completely before detaching relevant fitting and/or 

tube from the pressure port.

(4)Be careful not to mix foreign materials in the pressure port.

Mixing a foreign material in the port may cause product failure or air leakage. If there is a possibility of mixing foreign materials into the port, use 

the product with a filter.

(5)Be aware of possibilities of accidents related to the strap.

Swinging the product intentionally in the air by holding the strap may result in unexpected accidents, injury or product damage. Be sure to use the 

product with its main body gripped firmly.

(6)Be careful not to give external shocks to the products.

External shocks such as hitting and dropping may cause product failure.

(7)Installation of attachments

When installing an attachment to the pressure port, the tightening torque should be set at around 1 N・m. Make sure to use“insert ring”(attachment)

when you pour liquid into a tube connected with“one-touch fitting”(attachment).

(8)Attention for analog output function using.

No voltage or signal should be applied to the CAL terminal by tha faulty wiring. Failure to heed this warning may result in product malfunction or 

damage to the main body.

Moreover, do not lose the battery cover when the analog output function is used.

(9)Caution for handling of protect cover

The protect cover is made of silicon rubber. Never use fluids as media which are not compatible with the silicon rubber. The main purpose of the 

protect cover is to prevent liquid leak into the main body when the attachment and tube are detached from the product. It doesnʼt serve as drip-proof, 

so please wipe off in case any liquid leak takes place.

●The  product  is  not  explosion-proof.  Do  not  use  the  product  in  designated  areas  where 

explosion-proof specification is required.

●The product is not drip-proof /dust-proof. 
●The product is designed for measurement of intended articles and does not feature control 

functions that ensure security, such as prevention of accidents.

Instruction Manual

Ver.2.1

CAUTION

Head Office: Nishi-Shinjuku Prime Square, 

7-5-25 Nishi-Shinjuku, Shinjuku-ku, Tokyo 160-0023, Japan

https://www.nidec-copal-electronics.co

For inquiries regarding the product, please contact:

Summary of Contents for PG-100B Series

Page 1: ...ntionally in the air by holding the strap may result in unexpected accidents injury or product damage Be sure to use the product with its main body gripped firmly 6 Be careful not to give external shocks to the products External shocks such as hitting and dropping may cause product failure 7 Installation of attachments When installing an attachment to the pressure port the tightening torque should ...

Page 2: ... be careful not to allow the neutral detergent solution to enter the pressure port and or main body Infiltration of the neutral detergent may cause product failure 6 Caution for storage of protect cover Please store the protect cover in a cool and dark place if it is to be stored Ultraviolet rays exposure may make the quality of the cover deteriorate Please replace the protect cover with new one in...

Page 3: ...ero adjustment Hold Offset Auto power off Unit switching 0 5 F S 2digits Gases Liquids compatible with SUS316L 0 to 50 No freezing 15 to 85 RH or below No condensing 3 VDC AAA dry cell x 2 M5 female screw P 0 8 IP40 IEC standards 55 W 16 9 D 110 H Approx 100g Approx 1 000 hours 5 0 5V DC 5mA maximum External supply 1 000V 10mV Indicated value standard 102A Absolute 103R 102R 2 0 F S 2digits 1 0 F S ...

Page 4: ... function is invalid in 102A range because it is based on an absolute vacuum 3 Switching of auto power off function With the power turned off press the POWER button for one second or longer while pressing the HOLD button to set the mode to the switching mode of the auto power off function Choose the mode from four modes including P3 3 min P20 20 min PAt 20 min after the pressure reaches zero and PoF ...

Page 5: ...function or damage to the main body Please attach the protect cover to the main body from top side the side of pressure port to the end completely Be careful about the front side of the cover Strap canʼt be attached with the protect cover on the main body Periodical Calibration If you would like to calibration request please contact the sales office where you purchased Warranty The warranty period f...

Reviews: