10. TROUBLESHOOTING

According to the following table, review possilbe solutions for 
problems encountered. Do not disassemble the gauge by your-
self or attempt to repair. If you cannot resolve the fault yourself, 
please contact Nidec-Shimpo.

6

Failure

Possible Causes

Potential Solutions

Unit will not 
turn on

Low battery

Recharge and then re-
boot. If after 3-4 hours of 
charging time the battery 
does not properly hold a 
charge, the battery needs 
to be replaced. Contact 
Nidec-Shimpo.

No key 
sound

Key sound is 
turned off

Turn on the key sound in 
menu

No backlight Backlight is turned 

off

Turn on the backlight in 
menu

Error is too 
large

The gauge is not 
calibrated

Calibration of force gauge 
is required. After calibra-
tion if the error remains 
outside of the specifica-
tions, sensor may be 
damaged. Contact Nidec-
Shimpo to get RMA for 
return.

The RS232 serial port is used to connect the mini-printer to print 
the memory data stored on the gauge.

RS-232 Specifications:
-Hardware Flow Control: None
-Data word length: 8 bits
-Stop bit: 1bit
-Parity: None
-Baud rate: 38400

Fig. 8-2

8.2 Setpoint Output
Two NPN open collector setpoint outputs are available.

The internal circuit of the setpoint output is shown as Fig 8-2.

Pin7 with Pin6 will be connected when an overload alarm occurs.

In Preset Mode, Pin7 to Pin6 is connected when the measured 
value exceeds the upper limit. Pin4 to Pin6 is connected when the 
measured value passes below the lower limit.

CAUTION: Maximum permissible voltage: pin 7 to 6, pin 4 to 
6 must be lower than 35V ; pin 6 to 7, pin 6 to 4 must be lower 
than 6V .

Remember to remove the load after measurement. Applying a 
load for a long time period may affect the accuracy of the instru-
ment.

9. MAINTENANCE

9.1 Maintenance
After use, please keep the instrument body clean. Do not let oil 
and other substances persist on the body and screen so as not 
to damage the instrument. Remember to remove the load after 
measurement. Applying a load for a long time period may affect 
the accuracy of the instrument. 

9.2 Charging
When the battery is low, the icon “        ” will be displayed. The 
batteries should be charged immediately.

Connect the gauge and the charger with the USB cable. Then 
connect the charger with AC socket to start charging.

Summary of Contents for FG-3000R-S

Page 1: ...s Do not use damaged clamp Measure in line tension and compression forces only DO NOT attempt to measure forces at an angle to the measuring shaft damage to load cell and or shaft may result Do not attempt to repair or alter this instrument Warranty will be voided and damage to the unit may result Use and store within the stated temperature and humidity ranges or damage and failure may result When...

Page 2: ...PARATION 2 1 Confirm the model This series force gauge has 5 ranges available each model corre sponding to the capacity and resolution shown on the last page of this manual Select the appropriate model you need before use 3 SETUP 3 1 Measurement The Measurement menu contains the Unit of measure and Mea surement Mode sub menus as shown in Fig 3 1 Fig 3 1 3 2 Select Units The measuring units can be ...

Page 3: ...s the peak value shown on the display 4 1 Memory Fig 4 1 Memory menu contains three submenus Browse Print Delete all as shown in Fig 4 1 You can browse stored data or print all the data via the FG PRINTER mini printer sold separately You may also delete all the records in the Delete all sub menu 4 2 Browse In the Browse sub menu The data in memory can be reviewed in the order saved which is oldest...

Page 4: ... keypad un derneath while Reverse will allow the display to be up right with the keypad above Fig 5 1 b Fig 5 1 b 5 2 Auto Power The FG 3000R has an automatic power off function With Auto Power on if there is no operation performed within five minutes it will power off automatically Fig 5 2 Fig 5 2 5 3 Backlight The backlight can be set to turn on or off See Fig 5 3 Choosing the backlight to be of...

Page 5: ...after a period of usage It is recommended to send the force gauge to a specialized test ing organization for calibration If you have standard force weights or other standard load and a test stand you may utilize this function and procedure to calbrate the sensor Fig 5 6 a 1 Mount the force gauge 2 Use the tare by use of the ZERO key 3 Enter Calibration sub menu as in Fig 5 6 a The calibration inte...

Page 6: ... bits Stop bit 1bit Parity None Baud rate 38400 Fig 8 2 8 2 Setpoint Output Two NPN open collector setpoint outputs are available The internal circuit of the setpoint output is shown as Fig 8 2 Pin7 with Pin6 will be connected when an overload alarm occurs In Preset Mode Pin7 to Pin6 is connected when the measured value exceeds the upper limit Pin4 to Pin6 is connected when the measured value pass...

Page 7: ...ty 1000 1 100 220 Resolution 1 0 001 0 01 0 01 FG 3000R 5 Capacity 5000 5 500 1100 Resolution 1 0 001 0 1 1 FG 3000R 10 Capacity 10 1000 2250 Resolution 0 01 1 1 FG 3000R 20 Capacity 20 2000 4500 Resolution 0 01 1 1 FG 3000R 50 Capacity 50 0 5000 11 Resolution 0 01 1 0 01 ...

Page 8: ...SEALS USA Inc Instruments Division Tel 630 924 7138 Email support seals usa com ...

Reviews: