12. If calibration was unsuccessful, the display will show ERR.  Re-
move the calibration weight, then press UNIT. Repeat the proce-
dure to attempt for a correct calibration.

TROUBLESHOOTING

The following are general checkpoints. Please call Shimpo or your 
dealer for further assistance. 

Force Gauges does not turn on: Check that power adapter is con-
nected properly to unit and outlet.

Error codes are displayed: Turn unit off, then back on to reset. If 
error codes continue, see table below

CALIBRATION REqUIREMENTS

1. Secure and mount force gauge to a calibration stand and mount 
the gauge upside down.

2.  Obtain the appropriate full scale Metric range calibration weight 
of your unit.

3. Turn POWER OFF

4. Attach the fixture on the shaft of the force gauge. Thread and 
tighten onto shaft by hand only. 

5. Press and hold the UNIT, PEAK and ZERO switches.

6.  Press  and  release  the  POWER  switch  while  still  holding  the 
other 3 switches. Continue holding the 3 switches until the display 
reads CAL, now release the three switches.

7.  Press the UNIT switch. The display will blink SCN for 10 sec-
onds then show ZER. The force gauge is now ready for zero point 
calibration. 

8.  Press ZERO to zero calibrate.  Wait approximately 18 seconds. 
The  display  will  blink  SCN  then  PEK.  Do  not  press  any  other 
switches or move the sensing shaft during calibration.

9. Hang the calibration weight on the hook and stabilize. The pro-
cess  values  will  change.  The  gauge  is  now  ready  for  full-scale 
calibration.

10.  Press the PEAK switch to begin full scale calibration.  Display 
blinks  SCN.  Do  not  press  any  keys  or  touch  the  unit  or  weight 
during  calibration.  After  approximately  18  seconds,  the  display 
blinks END.  After an additional 5 seconds, the display blinks OK 
if successful.

11. Press the UNIT button, the unit automatically powers off.

5

CAL

-00-

“SCN”

-00-

ZER

88888

Err

88888

“SCN”

88888

“End”

88888

“oK”

88888

Small Display

Condition

Action

The load exceeds 
120% of its capacity

Remove excessive 
load.

EEPROM reading 
error

Turn off, then turn on 
again, if the display 
does not return to 
normal operation 
send unit in for repair.

EEPROM writing error

OVR

ERR

3

ERR

4

Summary of Contents for FGE-0.5XY

Page 1: ...pplications include wallboard wire pull wire crimps trig ger pull brake pedal aerospace ceramic springs food ripeness corrugated paper strength blister pack medical component muscle strength ergonomic testing adhesives syringe plunger needle sharpness and many more Series FGE XY Force Gauges Operation Manual NIDEC SHIMPO INSTRUMENTS Do not operate or store instrument in the following locations Exp...

Page 2: ...E PRODUCT DIAGRAM 2 ...

Page 3: ... high Units 0 29 7 mm Reversible by the push of a button Minus sign for tension Display Update 0 05 0 1 0 2 0 3 0 5 1 sec Sampling Rate 1000 Hz Overload 200 of F S 200 lb 150 F S Power Rechargeable Ni MH battery or AC adapter charger Auto Power Shut Off Yes not active if adapter charge is in use Battery Charge 10 hours when fully charged Recharge Time Approximately 17 hours when fully discharged T...

Page 4: ...asured with a minus sign to the left of the digits Note To eliminate the minus sign during tensile operation this may be changed Review the Function Setting section Compression Measurement To measure compression forces use the flat head attachment provided The display will not show any sign for compression Note To display a minus sign for compression forces consult the Function Settings section fo...

Page 5: ...til the display reads CAL now release the three switches 7 Press the UNIT switch The display will blink SCN for 10 sec onds then show ZER The force gauge is now ready for zero point calibration 8 Press ZERO to zero calibrate Wait approximately 18 seconds The display will blink SCN then PEK Do not press any other switches or move the sensing shaft during calibration 9 Hang the calibration weight on...

Page 6: ...DIMENSIONS ...

Reviews: