Nidec-Shimpo FGE-0.5XY Operation Manual Download Page 3

OPERATION

To Begin:

1.  Charge the batteries for approximately 17 hours before using  
the gauge. The batteries come discharged from the factory. 

2.  Hand-tighten appropriate attachment to units measuring 
shaft. Do not use wrench or any other device or tool to
tighten the attachments to the shaft.

3.  Press Power and release. The unit will display model name in  
the upper display and the capacity in the main display. 

Change Display Units

To change the display units, simply press UNIT and the units will 
change with each subsequent press of the button. 

Reverse the Display

If the gauge is used with the included hanger or mounted on a test 
stand, the display must be reversed.

1.  Press POWER and hold. The display will go blank.

2.  Press and hold UNIT.

3.  Release POWER while still continuing to hold down UNIT. Dis-
play will still remain blank.

SPECIFICATIONS

FGE-XY Model

FGE-0.5XY

FGE-1XY

FGE-2XY

FGE-5XY

FGE-10XY

FGE-20XY

FGE-50XY

FGE-100XY

FGE-200XY

Capacity

8 oz

0.5 lb

200.0 g

2.000 N

16 oz 

1 lb

500.0 g

5.000 N

32 oz

2 lb

1,000 g

10.00 N

5 lb

2.000 Kg

20.00 N

10 lb

5.000 Kg

50.00 N

20 lb

10.00 Kg

100.0 N

50 lb

20.00 Kg

200.0 N

100 lb

50.00 Kg

500.0 N

200 lb

100 Kg

1000N

Resolution

 0.01 oz
0.001 lb

0.1 g

0.001 N

0.01 oz

0.001 lb

0.1 g

0.001 N

0.01 oz 

0.001 lb

1 g

0.01 N

0.001 lb

0.001 Kg

0.01 N

0.01 lb

0.001 Kg

0.01 N

0.01 lb

0.01 Kg

0.1 N

0.01 lb

0.01 Kg

0.1 N

0.1 lb

0.01 Kg

0.1 N

0.1 lb

0.1 Kg

1N

Accuracy

±0.2% F.S.

Display

4-Digit LCD, 0.5” (12 mm) high, Units 0.29” (7 mm). Reversible by the push of a button. Minus sign for tension

Display Update

0.05, 0.1, 0.2, 0.3, 0.5, 1 sec

Sampling Rate

1000 Hz 

Overload

200% of F.S. (200 lb.: 150% F.S.)

Power

Rechargeable Ni-MH battery or AC adapter/charger

Auto Power Shut-Off

Yes (not active if adapter/charge is in use)

Battery Charge

10 hours (when fully charged)

Recharge Time

Approximately 17 hours when fully discharged

Temperature Range

32 - 104°F (0 - 40°C)

Dimensions

5.8 x 3 x 1.5” (147 x 75 x 38 mm) 

Product Weight

Approx. 1 lb  (450 g);  FGE-200XY  Approx. 1.1 lb  (500 g)

Package Weight

Approx. 2.25 lb ( 1.02 kg) for 0.5 lb and 1 lb ranges; 2.4 lb (1.09 kg) for 2 to 100 lb ranges;  2.75 lb (1.25  kg) for 200 lb range.

Certification

CE, RoHs

Included 
Accessories

AC adapter/charger, carrying case, hook, chisel, flat head, notched head, hanger, cone head, thread Adapter, extension rod

4.  Press POWER once more and release it while still holding UNIT.  
At this point the display will appear and be reversed. 

5.  Release UNIT; the display stays now in the reverse mode.

To return to the normal mode, repeat steps 1-5.

Select Average or Peak

The unit can display the live Averaged reading or the Peak read-
ing. Standard is Average.

If you want to measure the Peak force, follow this procedure:
1.  Press PEAK; the word PEAK will appear in the upper left corner 
of the display. If you wish to have a minus sign on the Peak, press 
PEAK again. 

2.  The display will freeze and show only the peak force on any 
given test.

3.    To  reset  the  Peak  reading,  press  ZERO  to  clear  the  current 
Peak reading and start over.

If  you  wish  to  go  back  to  Average  reading  mode,  press  PEAK 
again. The PEAK icon will disappear from the display.

3

Summary of Contents for FGE-0.5XY

Page 1: ...pplications include wallboard wire pull wire crimps trig ger pull brake pedal aerospace ceramic springs food ripeness corrugated paper strength blister pack medical component muscle strength ergonomic testing adhesives syringe plunger needle sharpness and many more Series FGE XY Force Gauges Operation Manual NIDEC SHIMPO INSTRUMENTS Do not operate or store instrument in the following locations Exp...

Page 2: ...E PRODUCT DIAGRAM 2 ...

Page 3: ... high Units 0 29 7 mm Reversible by the push of a button Minus sign for tension Display Update 0 05 0 1 0 2 0 3 0 5 1 sec Sampling Rate 1000 Hz Overload 200 of F S 200 lb 150 F S Power Rechargeable Ni MH battery or AC adapter charger Auto Power Shut Off Yes not active if adapter charge is in use Battery Charge 10 hours when fully charged Recharge Time Approximately 17 hours when fully discharged T...

Page 4: ...asured with a minus sign to the left of the digits Note To eliminate the minus sign during tensile operation this may be changed Review the Function Setting section Compression Measurement To measure compression forces use the flat head attachment provided The display will not show any sign for compression Note To display a minus sign for compression forces consult the Function Settings section fo...

Page 5: ...til the display reads CAL now release the three switches 7 Press the UNIT switch The display will blink SCN for 10 sec onds then show ZER The force gauge is now ready for zero point calibration 8 Press ZERO to zero calibrate Wait approximately 18 seconds The display will blink SCN then PEK Do not press any other switches or move the sensing shaft during calibration 9 Hang the calibration weight on...

Page 6: ...DIMENSIONS ...

Reviews: