Note:  To modify the auto power off function, consult the “Function 
Setting” section.

Tracking Function

The tracking function accounts and compensates for temperature 
drift.  It  is  advised  to  keep  this  function  running  for  optimal  per-
formance.  However,  when  measuring  very  minute  forces  (a  few 
ounces or grams) at a very slow rate, it may be advisable to dis-
able this function.

To turn the tracking function off, follow these steps:
1.  Turn POWER off

2. Press PEAK and UNIT switches simultaneously and hold.  Then 
press  and  release  POWER.  If  the  tracking  function  was  on,  the 
display will show TRK OFF momentarily to confirm successful dis-
able. 

3.  Release PEAK and UNIT

The  tracking  function  is  now  off.  Repeat  these  steps  to  activate 
the tracking function. The gauge displays TRK SET when reacti-
vated.

FUNCTION SETTING

To change the default parameters of the gauge, follow the proce-
dure to access the settings:

1.  Press and hold the ZERO button.

2.  Press and release the POWER button

3.    Each  time  the  PEAK  button  is  pressed,  the  unit  will  scroll 
through each of the functions.

4.  Press the ZERO button to exit.

Taking a Measurement

After the appropriate attachment has been hand-tightened on the 
instrument, tare the unit by pressing ZERO. 

Tension Measurement

To  measure  tensile  forces,  use  the  hook  attachment  provided.  
The display will show the force measured with a minus sign (-) to 
the left of the digits. 

Note:  To eliminate the minus (-) sign during tensile operation, this 
may be changed.  Review the “Function Setting” section.

Compression Measurement

To  measure  compression  forces,  use  the  flat  head  attachment 
provided.  The display will not show any sign for compression. 

Note:  To display a minus sign (-) for compression forces, consult 
the “Function Settings” section for making the change.

Low Battery

When  the  battery  charge  is  low,  LO  BAT  will  appear  indicating 
the batteries need to be charged.  To completely charge fully de-
pleted batteries requires approximately 17 hours of charging time 
with the unit off.  The units possess a smart charging system that 
shuts off the charging circuit once the batteries are fully charged. 

Note:  The unit may operate with the charging adapter while the 
batteries are being charged.  This will however lengthen charging 
time. 

Auto Power Off

If the gauge is on and there is no activity for 10 minutes, the unit 
automatically powers off to conserve the batteries.  At 9 minutes 
without activity, PWR appears to notify that there is 1 minute before 
auto power-off.  If the charging adapter is powering the gauge, the 
auto off function is inactive. 

4

F01

0001

F01

-0001

F01

0001

UNIT

UNIT

UNIT

UNIT

UNIT

UNIT

UNIT

UNIT

UNIT

UNIT

MINUS

PLUS

PEAK

Power OFF; 

PEAK

F02

3

F02

5

F02

10

F02

20

F02

1

F02

2

F02

3

F03

10

F03

OFF

F03

10

UNIT

UNIT

UNIT

Display Update 
Time

Plus or Minus
Sign

Auto Power
Off

Press and hold ZERO button. Press and release the POWER button.

ZERO; 
Save and Exit

Summary of Contents for FGE-0.5XY

Page 1: ...pplications include wallboard wire pull wire crimps trig ger pull brake pedal aerospace ceramic springs food ripeness corrugated paper strength blister pack medical component muscle strength ergonomic testing adhesives syringe plunger needle sharpness and many more Series FGE XY Force Gauges Operation Manual NIDEC SHIMPO INSTRUMENTS Do not operate or store instrument in the following locations Exp...

Page 2: ...E PRODUCT DIAGRAM 2 ...

Page 3: ... high Units 0 29 7 mm Reversible by the push of a button Minus sign for tension Display Update 0 05 0 1 0 2 0 3 0 5 1 sec Sampling Rate 1000 Hz Overload 200 of F S 200 lb 150 F S Power Rechargeable Ni MH battery or AC adapter charger Auto Power Shut Off Yes not active if adapter charge is in use Battery Charge 10 hours when fully charged Recharge Time Approximately 17 hours when fully discharged T...

Page 4: ...asured with a minus sign to the left of the digits Note To eliminate the minus sign during tensile operation this may be changed Review the Function Setting section Compression Measurement To measure compression forces use the flat head attachment provided The display will not show any sign for compression Note To display a minus sign for compression forces consult the Function Settings section fo...

Page 5: ...til the display reads CAL now release the three switches 7 Press the UNIT switch The display will blink SCN for 10 sec onds then show ZER The force gauge is now ready for zero point calibration 8 Press ZERO to zero calibrate Wait approximately 18 seconds The display will blink SCN then PEK Do not press any other switches or move the sensing shaft during calibration 9 Hang the calibration weight on...

Page 6: ...DIMENSIONS ...

Reviews: