![Nickel-Electro Clifton Range NE9-112S Manual Download Page 9](http://html1.mh-extra.com/html/nickel-electro/clifton-range-ne9-112s/clifton-range-ne9-112s_manual_1687522009.webp)
Operating Instructions
Before using the instrument, it is recommended to heat the drying chamber. Before the
heating process, please clean the drying chamber in order to avoid permanent stains. Turn
on the unit and set the temperature at maximum. Let the unit work for 3 hours. During the
heating, the unit is likely to produce a specific metallic smell.
Turning On
Switch the Clifton Range Drying Oven on using the mains switch located top right hand
side controls, the instrument will enter an auto diagnostics routine - if an error is found the
instrument stops or an appropriate screen is shown.
Once the diagnostic procedure concludes, the display shows an operation status screen and
measurement of temperature, which is calibrated to show the actual temperature in the
geometrical midpoint of the heating chamber.
The unit is inactive; nothing is running, display is just indicating chamber temperature with
date and time information.
CLIFTON RANGE DRYING OVEN NE9 SERIES
CLIFTON RANGE DRYING OVEN NE9 SERIES
9