28
3)
Enter 0.1 in the Sig field on the lower right to temporarily set the voltage of the Sig
detector to 0.1V.
4)
Take about 2.0 ng (of the standard gas) from the MB-1 and carry out measurement to
check the slope of the calibration curve (PEAK analysis) in LOW mode. If the slope of
the calibration curve is outside the allowable range of 6E+00 – 8E+00, change or enter
a D.P value of between 20 and 120 in the SPAN field of the Sig settings so that the slope
of the calibration curve falls within the allowable range.
* The D.P setup value is proportional to the slope of the calibration curve.