NI SCXI-1333 Installation Instructions

5. Expand the Number of Rows

The SCXI-1333 offers convenient methods for expanding the number 
of rows of a matrix using two or more SCXI-1333 terminal blocks.

Caution

When using matrix expansion plugs, only connect terminal blocks of the same 

type. Connecting different types of terminal blocks can damage the module or the 
equipment connected to the module. For example, you can connect two or more 
SCXI-1333 terminal blocks; however, do 

not

 connect the SCXI-1333 with any other 

type of terminal block.

Figure 4 shows how to use a matrix expansion plug with SCXI terminal 
blocks to expand the number of rows of a 4

×

32 matrix, as an example. 

The matrix expansion plug connects columns to expand the number 
of rows. To expand the number of rows of a matrix, connect a matrix 
expansion plug to the top or bottom column connectors of the two terminal 
blocks.

Figure 4.  

Expanding the Number of Rows with the Matrix Expansion Plug

1

Matrix Expansion Plug

2

Thumbscrews

3

Column Connectors

4

3

2

4

3

2

8

3

2

3

1

2

3

Mod

u

le 1

M

a

trix

Exp

a

n

s

ion

Pl

u

g

Mod

u

le 2

Summary of Contents for SCXI-1333

Page 1: ...s Spring terminals on the SCXI 1333 allow you to access each 4 16 matrix The SCXI 1333 provides terminals for scanner advanced output and external input trigger signals and contains two 128 pin DIN co...

Page 2: ...hould enter from the keyboard sections of code programming examples and syntax examples This font is also used for the proper names of disk drives paths directories programs subprograms subroutines de...

Page 3: ...voltages no greater than 150 V This module can withstand up to 800 V impulse voltage Do not use this module for connection to signals or for measurements within Categories II III or IV Do not connect...

Page 4: ...tab to secure the wire No bare wire should extend past the spring terminal Exposed wire increases the risk of a short circuit causing a failure 7 Connect the safety earth ground to the safety ground...

Page 5: ...XI 1333 Installation Instructions Figure 2 SCXI 1333 Parts Locator Diagram 1 Spring Terminals 2 Column Connectors 3 Rear Connector 4 Thumbscrew 5 Spring Pad 6 Strain Relief Screw 7 Strain Relief Bar 8...

Page 6: ...the SCXI 1129 if you have not already done so Refer to the NI Switches Getting Started Guide for more information 1 Plug the SCXI 1333 onto the front connector of the SCXI 1129 2 Tighten the top and...

Page 7: ...ect two or more SCXI 1333 terminal blocks however do not connect the SCXI 1333 with any other type of terminal block Figure 4 shows how to use a matrix expansion plug with SCXI terminal blocks to expa...

Page 8: ...imum current per channel 2 ADC 2 AAC Environmental Operating temperature 0 to 50 C Storage temperature 20 to 70 C Humidity 10 to 90 RH noncondensing Pollution Degree 2 Approved at altitudes up to 2 00...

Page 9: ...number or product line and click the appropriate link in the Certification column Environmental Management National Instruments is committed to designing and manufacturing products in an environmental...

Page 10: ...al Instruments trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products refer to the a...

Reviews: