4. Install the Terminal Block

To connect the SCXI-1333 to the SCXI-1129 front panel, refer to Figure 3 
and complete the following steps:

Note

Install the SCXI-1129 if you have not already done so. Refer to the 

NI Switches 

Getting Started Guide

 for more information.

1.

Plug the SCXI-1333 onto the front connector of the SCXI-1129.

2.

Tighten the top and bottom thumbscrews on the terminal block rear 
panel to hold it securely in place.

Figure 3.  

Installing the SCXI-1333 Terminal Block

1

Thumbscrews

2

Front Connector

3

SCXI-1129

4

SCXI-1333

4

1

3

2

Summary of Contents for SCXI-1333

Page 1: ...s Spring terminals on the SCXI 1333 allow you to access each 4 16 matrix The SCXI 1333 provides terminals for scanner advanced output and external input trigger signals and contains two 128 pin DIN co...

Page 2: ...hould enter from the keyboard sections of code programming examples and syntax examples This font is also used for the proper names of disk drives paths directories programs subprograms subroutines de...

Page 3: ...voltages no greater than 150 V This module can withstand up to 800 V impulse voltage Do not use this module for connection to signals or for measurements within Categories II III or IV Do not connect...

Page 4: ...tab to secure the wire No bare wire should extend past the spring terminal Exposed wire increases the risk of a short circuit causing a failure 7 Connect the safety earth ground to the safety ground...

Page 5: ...XI 1333 Installation Instructions Figure 2 SCXI 1333 Parts Locator Diagram 1 Spring Terminals 2 Column Connectors 3 Rear Connector 4 Thumbscrew 5 Spring Pad 6 Strain Relief Screw 7 Strain Relief Bar 8...

Page 6: ...the SCXI 1129 if you have not already done so Refer to the NI Switches Getting Started Guide for more information 1 Plug the SCXI 1333 onto the front connector of the SCXI 1129 2 Tighten the top and...

Page 7: ...ect two or more SCXI 1333 terminal blocks however do not connect the SCXI 1333 with any other type of terminal block Figure 4 shows how to use a matrix expansion plug with SCXI terminal blocks to expa...

Page 8: ...imum current per channel 2 ADC 2 AAC Environmental Operating temperature 0 to 50 C Storage temperature 20 to 70 C Humidity 10 to 90 RH noncondensing Pollution Degree 2 Approved at altitudes up to 2 00...

Page 9: ...number or product line and click the appropriate link in the Certification column Environmental Management National Instruments is committed to designing and manufacturing products in an environmental...

Page 10: ...al Instruments trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products refer to the a...

Reviews: