© National Instruments
|
A-3
Operating Environment
NI PXIe-8135
Ambient temperature range
Standard ............................................ 5 to 50 °C
1
(Tested in accordance with IEC-60068-2-1 and
IEC-60068-2-2. Meets MIL-PRF-28800F
Class 3 high temperature limit.)
Extended Temperature
Option ............................................... 0 to 55 °C
1
(Tested in accordance with IEC-60068-2-1 and
IEC-60068-2-2. Meets MIL-PRF-28800F
Class 3 low temperature limit and
MIL-PRF-28800F Class 2 high temperature
limit.)
Relative humidity range............................ 10% to 90%, noncondensing
(Tested in accordance with IEC-60068-2-56.)
Caution
The operating temperature must not be exceeded, even when used in a
chassis with a higher temperature range.
Storage Environment
NI PXIe-8135
Ambient temperature range
Standard ............................................ -40 to 65 °C
(Tested in accordance with IEC-60068-2-1 and
IEC-60068-2-2. Meets MIL-PRF-28800F
Class 3 low temperature limit.)
Extended Temperature
Option ............................................... -40 to 71 °C
(Tested in accordance with IEC-60068-2-1 and
IEC-60068-2-2. Meets MIL-PRF-28800F
Class 3 limits.)
Relative humidity range............................ 5% to 95%, noncondensing
(Tested in accordance with IEC-60068-2-56.)
Shock and Vibration
Operating Shock ............................................... 30 g peak, half-sine, 11 ms pulse
(Tested in accordance with IEC-60068-2-27.
Meets MIL-PRF-28800F Class 2 limits.)
1
Processor should not throttle CPU frequency under reasonable, worst case processor work loads in high
operating temperatures.