NI PCI-5122 Getting Started Manual Download Page 7

4.

Touch any metal part of the computer to discharge any static electricity.

5.

Insert the module into the slot you selected. Gently rock the module in to place without
forcing it.

Figure 3. Module Installation

1

2

3

1. Module
2. System Expansion Slot
3. PC

6.

Secure the module mounting bracket to the computer back panel rail.

Note

 

Tightening the top mounting screw increases mechanical stability and

also electrically connects the front panel to the chassis, which can improve the
signal quality and electromagnetic performance.

7.

Some computer manufacturers use a securing lever made of plastic to secure PCI devices;
such a lever is unacceptable and must be removed. Use the screw provided in the kit to
screw down the device. Otherwise, you must use a different computer chassis.

8.

Replace any access panels on the computer case.

9.

Power on your computer.

10. Verify that spread-spectrum clocking is enabled in the PC BIOS. For information about

how to verify this setting, refer to the PC user documentation

Note

 

For SMC-based devices, spread-spectrum clocking varies the clock

signal to spread the timing clock signal over a small frequency range. Disabling
spread-spectrum clocking may affect the accuracy of device specifications.

NI PXI/PXIe/PCI-5122 Getting Started Guide

  | 

© National Instruments

  | 

7

Summary of Contents for PCI-5122

Page 1: ...netic Compatibility Guidelines 2 Verifying the System Requirements 2 Unpacking the Kit 2 Kit Contents 3 Preparing the Environment 3 PXI Modules 3 PXI Express Modules 4 PCI Modules 4 Installing the Software 4 Installing the PXI PXIe 5122 5 Installing the PCI 5122 6 Front Panel Connectors and Indicators 8 Configuring the PXI PXIe PCI 5122 in MAX 9 Programming the PXI PXIe PCI 5122 10 NI SCOPE Exampl...

Page 2: ...urthermore any changes or modifications to the product not expressly approved by National Instruments could void your authority to operate it under your local regulatory rules Verifying the System Requirements To use the PXI PXIe PCI 5122 your system must meet certain requirements For more information about minimum system requirements recommended system and supported application development enviro...

Page 3: ...ng the PXI PXIe PCI 5122 meets the following specifications PXI Modules Operating Environment Ambient temperature range 0 C to 55 C Tested in accordance with IEC 60068 2 1 and IEC 60068 2 2 Relative humidity range 10 to 90 noncondensing Tested in accordance with IEC 60068 2 56 Maximum altitude 2 000 m at 25 C ambient temperature Pollution Degree 2 Indoor use only NI PXI PXIe PCI 5122 Getting Start...

Page 4: ...nd IEC 60068 2 2 Relative humidity range 10 to 90 noncondensing Tested in accordance with IEC 60068 2 56 Maximum altitude 2 000 m at 25 C ambient temperature Pollution Degree 2 Indoor use only Caution Clean the hardware with a soft nonmetallic brush Make sure that the hardware is completely dry and free from contaminants before returning it to service For complete specifications refer to specifica...

Page 5: ...or to installation Do not install a module if the backplane is damaged 4 Remove the black plastic covers from all the captive screws on the module front panel 5 Identify a supported slot in the chassis The following figure shows the symbols that indicate the slot types Figure 1 Chassis Compatibility Symbols NI PXIe 1062Q 1 2 3 4 5 1 PXI Express System Controller Slot 2 PXI Peripheral Slot 3 PXI Ex...

Page 6: ...ze cooling air flow depending on your application 12 Power on the chassis Note For more information about maximizing cooling air flow refer to the Maintain Forced Air Cooling Note to Users included in your kit Related Information Why Is the ACCESS LED Off When the Chassis Is On on page 12 Installing the PCI 5122 1 Power off and unplug the computer 2 Access the computer system expansion slots This ...

Page 7: ...facturers use a securing lever made of plastic to secure PCI devices such a lever is unacceptable and must be removed Use the screw provided in the kit to screw down the device Otherwise you must use a different computer chassis 8 Replace any access panels on the computer case 9 Power on your computer 10 Verify that spread spectrum clocking is enabled in the PC BIOS For information about how to ve...

Page 8: ...tors Label Function Connector Type CH 0 Analog input BNC female CH 1 Analog input BNC female TRIG External analog trigger BNC female CLK IN Sample clock input and Reference clock input SMB jack CLK OUT Sample clock output and Reference clock output SMB jack AUX I O PFI 0 and PFI 1 lines for digital trigger input output and probe compensation PFI 1 only 9 pin mini circular DIN 8 ni com NI PXI PXIe ...

Page 9: ...X is automatically installed with NI SCOPE 1 Launch MAX 2 In the configuration tree expand Devices and Interfaces to see the list of installed devices Installed devices appear under the name of their associated chassis 3 Expand your Chassis tree item MAX lists all devices installed in the chassis Your default device names may vary Note If you do not see your device listed press F5 to refresh the l...

Page 10: ...rol waveforms directly from your computer You can also run multiple sessions of the NI SCOPE SFP simultaneously NI SCOPE Instrument Driver LabVIEW Available on the LabVIEW Functions palette at Measurement I O NI SCOPE The NI SCOPE API configures and operates the device hardware and provides customizable acquisition control analysis and measurement options using LabVIEW VIs or LabWindows CVI functi...

Page 11: ...sic Locate examples in the file directory Users Public Public Documents National Instruments NI SCOPE examples Making a Measurement You can make a measurement interactively using the NI SCOPE SFP or programmatically using LabVIEW Making a Measurement with NI SCOPE SFP 1 Connect CH 0 to an input signal 2 Launch the NI SCOPE SFP from the Start menu 3 In the Select Device dialog box select the device...

Page 12: ...ble between the PCI devices to physically share triggers and or clocks 2 Launch MAX 3 In the MAX configuration pane right click Devices and Interfaces and select Create New 4 In the dialog box that opens select NI RTSI Cable 5 Click Finish 6 In the MAX configuration pane right click the NI RTSI cable select Add Device to NI RTSI Cable and select the device you want to add For more information abou...

Page 13: ...PXI PXIe PCI 5122 Doesn t Appear in MAX 1 In the MAX configuration tree expand Devices and Interfaces 2 Expand the Chassis tree to see the list of installed hardware and press F5 to refresh the list 3 If the module is still not listed power off the system ensure that all hardware is correctly installed and restart the system 4 Navigate to the Device Manager Operating System Description Windows 10 ...

Page 14: ...2 Reinstall the device and make any necessary adjustments to make sure that the device is effectively cooled 3 Power on the computer or chassis Note The thermal shutdown error is reported until the device has cooled to an acceptable operating temperature and has been successfully reset For more information about cooling the device refer to the Maintain Forced Air Cooling Note to Users included in ...

Page 15: ... MHz Oscilloscope NI PXIe 5160 Worldwide Support and Services The NI website is your complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Application Engineers Visit ni com services for NI Factory Installation Services repairs extended warranty and other ser...

Page 16: ...erein are trademarks or trade names of their respective companies For patents covering NI products technology refer to the appropriate location Help Patents in your software the patents txt file on your media or the National Instruments Patent Notice at ni com patents You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI produ...

Reviews: