Table 2.
Equipment Required for Calibrating the NI 5421/5441
Instrument
Recommended Instrument
Applicable Specification
Minimum Specifications
Digital multimeter (DMM)
NI PXI-4070
Agilent/HP 34401A
Keithley 2000
AC accuracy, DC gain and
offset, and frequency accuracy
DCV accuracy:
≤
0.05%
DC input impedance:
≥
1 G
Ω
ACV accuracy:
≤
0.16%
AC input impedance:
≥
1 M
Ω
Bandwidth:
≥
100 kHz
Male banana-to-female BNC adapter
—
—
Male BNC-to-female SMB cable
—
50
Ω
, RG-223
Frequency counter
Agilent/HP 53131A or
HP 53132A with timebase option
001, 010, or 012
Frequency accuracy
Ability to measure 10 MHz or higher
sine waves
Frequency accuracy to ±500 ppb
Male BNC-to-female SMB cable
—
50
Ω
, RG-223
Power meter/sensor
R&S NRP-Z91
Frequency response (flatness)
VSWR: (50 kHz to 120 MHz)
≤
1.11
Relative power accuracy:
≤
0.022 dB
Type N female-to-SMB plug adapter
Pasternak PE9316
VSWR: 1.3