![NI 5421 Calibration Procedure Download Page 20](http://html1.mh-extra.com/html/ni/5421/5421_calibration-procedure_1685638020.webp)
20
|
ni.com
|
NI 5421/5441 Calibration Procedure
3.
Call
niFgen_SetAttributeViReal64
to set the gain (NI-FGEN Gain property) using
the following parameters:
•
vi
: The session handle returned from
niFgen_init
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_ARB_GAIN
•
value
: The
Gain
value listed in Table 6 for the current iteration
4.
Call
niFgen_SetAttributeViInt32
to choose the positive full-scale DC waveform
(NI-FGEN Arbitrary Waveform Handle property) using the following parameters:
•
vi
: The session handle returned from
niFgen_init
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_ARB_WAVEFORM_HANDLE
•
value
: The
wfmHandle
Verifying the Main Analog Path Gain
section (positive full-scale handle)
5.
Call
niFgen_InitiateGeneration
(niFgen Initiate Generation VI) to initiate the
waveform generation using the following parameter:
•
vi
: The session handle returned from
niFgen_init
6.
Measure the positive DC voltage from the NI 5421/5441. This value is the
Measured
Positive Full-Scale Value
.
7.
Call
niFgen_AbortGeneration
(niFgen Abort Generation VI) to abort the waveform
generation using the following parameter:
•
vi
: The session handle returned from
niFgen_init
8.
Call
niFgen_SetAttributeViInt32
to choose the negative full-scale DC waveform
(NI-FGEN Arbitrary Waveform Handle property) using the following parameters:
•
vi
: The session handle returned from
niFgen_init
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_ARB_WAVEFORM_HANDLE
6
0.750000
0.750000
- 0.750000
±0.025000
±0.003000
±0.001200
7
0.710000
0.710000
- 0.710000
±0.025000
±0.002840
±0.001136
Note
:
Offset
= ((
Measured Positive Full-Scale Value
) + (
Measured Negative Full-Scale Value
))/2
Error Positive Full-Scale Value
= (
Measured Positive Full-Scale Value
) -
Offset
- (
Ideal Positive
Full-Scale Value
)
Error Negative Full-Scale Value
= (
Measured Negative Full-Scale Value
) -
Offset
- (
Ideal Negative
Full-Scale Value
)
Table 6.
Values for Verifying the Gain of the Direct Analog Path (Continued)
Iteration
Gain
Ideal
Positive
Full-Scale
(Volts)
Ideal
Negative
Full-Scale
(Volts)
Offset
Limit
(Volts)
As-Found
Limits
(Volts)
As-Left
Limits
(Volts)