NI 5421/5441 Calibration Procedure
|
©
National
Instruments
|
39
•
attributeID
:
NIFGEN_ATTR_POST_AMPLIFIER_ATTENUATION
•
value
:
0
3.
Call
niFgen_SetAttributeViInt32
to set the offset DAC value (NI-FGEN Offset
DAC Value property) using the following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_OFFSET_DAC_VALUE
•
value
:
32000
Repeat steps 4 through 7 for each of the 10 iterations listed in Table 11, changing the
Analog
Filter Enable
,
Preamplifier Attenuation
, and
Current Configuration
values for each iteration.
4.
Call
niFgen_SetAttributeViBoolean
to set the analog filter state (NI-FGEN
Analog Filter Enabled property) using the following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
Table 11.
Attributes and Values for Main Analog Path Preamplifier Gain
Iteration
Analog Filter
Enable
Preamplifier
Attenuation
Current Configuration
1
VI_FALSE
0
NIFGEN_VAL_CAL_CONFIG_MAIN_
PATH_FILTER_OFF_0DB
2
VI_FALSE
3
NIFGEN_VAL_CAL_CONFIG_MAIN_
PATH_FILTER_OFF_3DB
3
VI_FALSE
6
NIFGEN_VAL_CAL_CONFIG_MAIN_
PATH_FILTER_OFF_6DB
4
VI_FALSE
9
NIFGEN_VAL_CAL_CONFIG_MAIN_
PATH_FILTER_OFF_9DB
5
VI_FALSE
12
NIFGEN_VAL_CAL_CONFIG_MAIN_P
ATH_FILTER_OFF_12DB
6
VI_TRUE
0
NIFGEN_VAL_CAL_CONFIG_MAIN_
PATH_FILTER_ON_0DB
7
VI_TRUE
3
NIFGEN_VAL_CAL_CONFIG_MAIN_
PATH_FILTER_ON_3DB
8
VI_TRUE
6
NIFGEN_VAL_CAL_CONFIG_MAIN_
PATH_FILTER_ON_6DB
9
VI_TRUE
9
NIFGEN_VAL_CAL_CONFIG_MAIN_
PATH_FILTER_ON_9DB
10
VI_TRUE
12
NIFGEN_VAL_CAL_CONFIG_MAIN_
PATH_FILTER_ON_12DB