6-2
WF1943B
6.1 Error message
Self-check is conducted at power on and in case of an abnormality, an error message is displayed. An error
message is also displayed if an erroneous operation is conducted.
Error message contents, causes and corrective measures are indicated in the following tables.
Power on error
Error message
Cause
Corrective measures
BACKUP MEMORY LOST
Battery backup
memory contents
destroyed.
Backup battery probably depleted.
Contact dealer.
Press the ENTER key to start the system at
the factory settings.
CALIBRATION MEMORY LOST
Calibration data
destroyed.
Contact dealer. Although possible to start
by pressing the ENTER key, accuracy
cannot be guaranteed.
SYSTEM TEST FAILED 001
Internal ROM sum
check error.
Contact dealer.
SYSTEM TEST FAILED 002
Internal RAM
read/write error.
Contact dealer.