6.4 Performance Testing
HSA42011
6-8
6.4.3 Measurement of frequency characteristics
Connection
Connect a frequency response analyzer (FRA) and terminating resistor as shown in "
Figure 6-4
Measurement of Frequency Characteristics
".
Setting
Set this product as follows:
Item Setting
Input A
Input impedance
50
Bias addition
OFF, dial 5.00
Polarity switching (INVT)
OFF
Gain setting
50 (CAL)
Terminating resistor
50
Set the FRA as follows:
Item Setting
Output
Sine wave 0.565 Vpeak
Sweep frequency
100 Hz to 1 MHz, log sweep
Analysis ch1/ch2
Indication
logF
logR
Test procedure
<1> Turn on the outputs of this product and FRA.
<2> Perform UP (or DOWN) sweep operation for measurement at 100 Hz to 1 MHz.
<3> Move the cursor after measurement to read the gains at 400 Hz, 100
kHz, 500 kHz, and 1 MHz.
Summary of Contents for HSA42011