6.4 Performance Testing
HSA42011
6-6
6.4.1 Measurement of maximum output voltage (AC)
Connection
Connect a signal generator, AC voltmeter, oscilloscope, and terminating resistor as shown in "
Figure
6-2 Measurement of Maximum Output Voltage
".
Setting
Set this product as follows:
Item Setting
Input A
Input impedance
50
Bias addition
OFF, dial 5.00
Polarity switching (INVT)
OFF
Gain setting
50 (CAL)
Terminating resistor
50
Test procedure
<1> Set the waveform of the signal generator to sine wave, and set the frequencies shown in "
Table
6-1 Judgment of HSA42011 Performance
".
<2> Set the output on/off switch of this product to ON.
<3> Increase the output voltage of the signal generator gradually from 0 V.
<4> Observe the waveform on the oscilloscope. Record the output voltage value when the waveform
begins to be distorted or the overload LED (OVLD) begins to be glow.
本製品
WF シリーズ
1CH 30 MHz
信号発生器
オシ ロスコープ
10:1プローブ
終端抵抗
電圧計
Figure 6-2. Measurement of Maximum Output Voltage (AC)
WF series
1CH 30MHz
Signal generato
r
This product
AC voltmeter
Oscilloscope
10:1 probe
Terminating
resistor