PPA45xx Quick User Guide
Page 36
Three Phase Three Wattmeter– Delta Connections
To configure PPA45xx to calculate the correct phase power when the Load
topology is in a Delta Configuration;
Access “Power Analyzer” mode either through the “Mode” or the “POWER” button
as per the screenshot below
Press 8 times until red box surrounds “conversion”
Press to open up dropdown menu selections
Press until red box surrounds “delta – star”
Press “ENTER” to confirm selection
Three
Phase
Source
or
A
CH1
A
CH2
A
CH3
Three
Phase
Load
or
V
CH1
V
CH2
V
CH3
HI
HI
HI
HI
HI
HI
LO
LO
LO
Ph1
Ph1
Ph2
Ph3
Ph3
Ph2
LO
LO
LO