16.3 STANDARD MODE
Most normal communication with the MS260i, especially communication handled inside closed
application software, is done in Standard Mode. This mode follows all of the rules on the previous
page, plus there are some additional features for error handling.
The MS260i stores errors internally in a status byte and an error byte, where each bit has a specific
meaning. Details of the status and error bytes can be found in this manual under Error Codes.
In all cases, Statements must be terminated properly with [lf] for MS260i to take any action.
Statements that are not understood or parameters that are unacceptable generate an error code in
the status byte, i.e. status byte not equal to 0. Moreover, the status byte must be explicitly read. It
is not returned after general Statements.
Correct Statements are handled as follows:
Commands
Under RS-232, Statements are echoed back to the PC. Thus, after sending a Command you
should read back this echo. There is no further response to a Command.
Example:
Send: GOWAVE 500[lf]
Response: GOWAVE 500[lf]
Under IEEE-488 and USB, there is no echo. There is no automatic response to a Command.
Example:
Send: GOWAVE 500[lf]
Response: {none}
Queries
Under RS232, Statements are echoed back to the PC. Additionally, there will be an explicit
response to that query. Thus, after sending a Query the echo and the answer should be read back
together.
Example:
Send : WAVE?[lf]
Response: WAVE?[lf]500[cr][lf]
Under IEEE-488 and USB, there is no echo. Thus, the Response can be read back by itself.
Example:
Send : WAVE?[lf]