OPS-A
Arc Lamp Power Supplies
Page 21
Note. If a lamp that was not purchased from Newport Corporation is being used, use the
manufacturer’s current specifications for operation. If the lamp operating power and voltage are
known but the operating current is not, determine the current setting by using Ohm’s Law:
Amperes = Power ÷ Voltage
4. Press CLEAR to clear the vertical menu from the screen and observe the desired operating parameter
in real time while the lamp is being operated. The lamp is now ready for ignition.
Please note that the lamp operation mode cannot be changed while the lamp is being operated. If the
operating mode is to be changed, power off the lamp and change the lamp operation mode before
reigniting the lamp.