New Era Pump Systems Inc.
www.SyringePump.com
Model NE-1000 Multi-Phaser™
Publication #1200-01
45
04/06/16
Stop Only:
Falling
edge Stops the Pumping Program. This configuration only allows the
stopping/pausing of the Pumping Program. This would be useful, for example, with an end
of travel limit switch. Also, this switch can be used as a power on homing switch.
Stop Only Reversed:
Same as ‘Stop Only’, but operates on the
Rising
edge to stop the Pumping Program.
Run on Low Level:
Starts the pump whenever the
level is low
Run on High Level:
Starts the pump whenever the
level is high
Stop on Low Level:
Stop the pump whenever the
level is low
Stop on High Level:
Stop the pump whenever the
level is high
Trigger off:
Trigger control is disabled
Event Trap Trigger:
Redirects a Stop Operational Trigger to the Event Trap.
Button Event Trigger:
Redirects ‘Stop’ key to the Event Trap.
Pump Motor Operating (Pin 7):
This output provides an external signal indicating when the pump motor is
operating. This pin is configured with the
n
setup command, or the "ROM" remote command.
When set to 0, the output is only at logic high when the motor is operating (pumping). When set to 1, the
output is logic high when the motor is operating or when the Pumping program is executing a pause timer.
Otherwise, the output is a logic low.
Pumping Direction Controls (Input: Pin 3; Output: Pin 8):
Allows bi-directional control of the pumping
direction. The input pin, when activated, emulates the pumping direction key, changing the pumping direction.
This function, therefore, is only applicable where the pumping direction key would be applicable. The
function of the input pin is configured with the
aa
setup command, or the "DIN" remote command.
When the mode is set to reciprocating pumps ("rE" setup command), then if the current pumping direction is
withdraw, a falling edge sets the direction to infuse. If the current pumping direction is infuse, a rising edge
sets the direction to withdraw. Otherwise, this input pin has no effect.
When the mode is set to dual pumps ("dU" setup command), then if the current pumping direction is withdraw,
a rising edge sets the direction to infuse. If the current pumping direction is infuse, a falling edge sets the
direction to withdraw. Otherwise, this input pin has no effect.
Dual and reciprocating pumping systems are created using 2 pumps attached with the accessory cable CBL-
TTL-1.
The output pin provides an output signal to external devices indicating the direction of pumping. A logic low
indicates withdraw, and a logic high indicates infuse. For example, this pin can be used to control an external
valve, allowing the syringe to refill from a reservoir.
11.2
TTL I/O Control from the Pumping Program
Various Pumping Program functions can define how the pump reacts to levels on the TTL I/O connector or set
output levels. These are summarized in the following table:
Pumping Program
Function
TTL I/O Control Pin
Pin
#
Action
OUT.n
Program Output
5
Set logic level output to ‘n’
EV:nn
Event Trigger
4
Falling edge triggers a jump to Phase ‘nn’
ES:nn
Event Square wave Trigger
4
Rising or falling edge triggers a jump to Phase 'nn'
IF:nn
Program Input
6
Low level causes a jump to Phase ‘nn’
PS:00
Operational Trigger
2
Trigger activation resumes Program operation
11.3
TTL I/O Control from RS-232
The logic levels of pins 2, 3, 4, and 6 can be queried from an attached computer using the RS-232 ‘IN’
command.
The output logic level of pin 5 can be set with the RS-232 ‘OUT’ command.
50