TTL I/O Operational Trigger Default Configuration
aa
Configures the default mode of the TTL I/O ‘Operational Trigger’ (pin 2) that will control the
Pumping Program’s operation. This default setting can be overridden within a Pumping Program. (See sec.
11.1,
TTL I/O Operational Controls
). The 2 letter configuration Setting parameter to the right of the colon (:)
is defined in column ‘Setting’ as follows:
Setting
<aa>
RS-232 Command Setting
FUN TRG <n>
<n> =>
Name
Function
Ft
0
Foot Switch
Falling edge starts or stops the Pumping Program
FH
1
Foot Switch Hold
Falling edge starts the Pumping Program
Rising edge stops the Pumping Program
F2
2
Foot Switch Reversed
Rising edge starts or stops the Pumping Program
LE
3
Level Control
Falling edge stops the Pumping Program
Rising edge starts the Pumping Program
St
4
Start Only
Falling edge starts the Pumping Program
t2
5
Start Only Reversed
Rising edge starts the Pumping Program
SP
6
Stop Only
Falling edge stops the Pumping Program
P2
7
Stop Only Reversed
Rising edge stops the Pumping Program
rL
8
Start on Low Level
Start Pumping Program on a low level
rH
9
Start on High Level
Start Pumping Program on a high level
SL
10
Stop on Low Level
Stop Pumping Program on a low level
SH
11
Stop on High Level
Stop Pumping Program on a high level
oF
12
Trigger off
Disable trigger
Et
13
Trigger Event
Pumping Program Function Only: Stop Trigger
triggers an event instead of stopping the pump
Bt
14
Button Trigger Event
Pumping Program Function Only: Pressing ‘Stop’
key triggers an event instead of stopping the pump
8.5
TTL I/O Directional Control Input Configuration
aa
Configures how the TTL input ‘Pumping Direction’ (pin 3) will control the pumping direction.
(See sec. 11.1,
TTL I/O Operational Controls
). The 2 letter configuration parameter to the right of the colon
(:) is defined as follows:
Setting Name
Function
rE
Reciprocating Pumps
Falling edge: Infuse; Rising edge: Withdraw
dU
Dual Pump
Falling edge: Withdraw; Rising edge: Infuse
The setting names are relevant to a 2 pump system, whereby the ‘Directional Control Input’ TTL pin is
attached to the second pump’s ‘Pumping Direction Output’ TTL pin.
In addition, “Directional Control Input” will affect how the “Sticky Direction” function operates. See “ Sticky
Direction”, sec: 6.8.1