
New Era Pump Systems Inc.
www.SyringePump.com
Model NE-1000 Multi-Phaser™
Publication #1200-01
39
04/06/16
Query response: { INF | WDR | STK}
While pumping, when set to “Sticky Direction”, query will respond with the current pumping
direction ( INF or WDR).
10.4.2
Pump Operational Commands
START PUMPING PROGRAM
RUN [ <phase data> ]
[E [<phase data> ] ]
Starts the Pumping Program operation.
If the Pumping Program was paused, then the Pumping Program resumes at the point where it was
stopped. Otherwise, the Pumping Program starts from Phase 1.
Sub-Programs: If a Phase number is specified (<phase data>), then the Pumping Program will start at
the specified Phase number. By programming sub-programs in different sections of the Pumping
Program memory, this command can be used to individually execute different sub-programs.
The pump cannot be started if an alarm condition exists.
E [ <phase data> ]
Trigger a Pumping Program Event while the Pumping Program is active.
Triggers a pre-defined event defined with the Pumping Program’s Event function, causing an
immediate jump to the Pumping Program Phase defined by the event function.
If <phase data> is specified, the program will immediately jump to the Pumping Program Phase
specified by <phase data>, and cancel any other event set by the Pumping Program.
PURGE PUMP
PUR
Starts purge. Pump infuses or withdraws at the top speed, depending on the pumping direction.
STP command or ‘Start/Stop’ key will stop the pump.
STOP PUMPING PROGRAM
STP
If the Pumping Program is operating, the pump will be stopped and the Pumping Program will be
paused.
If the Pumping Program is paused, the stop command will cancel the pause and reset the Pumping
Program to Phase 1.
VOLUME DISPENSED
DIS
Queries volume dispensed only. Set not applicable.
Response:
I <float> W <float> <volume units>
Where: “I <float>“ refers to the infusion volume dispensed, and “W <float>” refers to the withdrawn
volume.
CLEAR VOLUME DISPENSED
CLD { INF | WDR }
Sets the Infused or withdrawn volume dispensed to 0. Command is only valid while the Pumping
Program is not operating. Query is not applicable.
INF
= Infusion volume
WDR
= Withdrawn volume
Query is not applicable.
10.4.3
Configuration and Setup Commands
New settings for any of the following commands will be stored in the non-volatile memory.
LOW MOTOR NOISE
LN [ <on-off> ]
Set/query low motor noise mode. Increases motor stepping resolution.
44
44 (of 55 )