Compass calibration (CI model only)
If the System was exposed to a strong magnetic field or
was not in use for a long time, the compass accuracy may
diminish.
To return the compass into working condition:
Select
CALc
from
CAL
menu
Keeping the unit horizontally, perform one full
rotation around vertical axis within half a minute and
then one full rotation around vertical axis with device
upside down during half a minute.
If one rotation does not restore calibration – perform
two rotations.
The normal compass functioning will be restored.
Setting interface format (CI model only)
In this mode data communication standard is set. The unit
can output measurement data via RS-232 port in either
PC format (description of the proprietary format can be
obtained from Newcon upon a request) or in the format
accepted by PLGR / DALGR.
To select required data format:
Select
PLGr
or PC from
SEt
menu
Pressing Action button will activate PC or PLGR
format, the display name of format that can be
activated will be shown.