- 28 -
5. Replacement of Consumables
Battery Replacement
The following shows the remaining battery level indicator. When a battery is running out,
it is unable to detect gas, as “E-B” appears with intermittent buzzer sound. (LCD will be
off when the battery is dead)
1) Press and hold the [ POWER] button to turn OFF the power.
2) Loosen the battery cover by turning it counterclockwise
with a cross slot screwdriver with point size 2, and
remove the battery cover.
3) Pull the battery cover upward with fingers.
WARNING
MENU
Battery Cover
•
Make sure to use a screwdriver with
point size 2 (see on page 32 for the
shapes). If the screwdriver is too small,
the battery cover can be damaged.
•
Turn the screwdriver slowly. The battery
cover will be damaged if it is turned by
force.
NOTE
•
Make sure to use a new battery.
•
Low temperature may shorten the battery life due to the battery property. If
the remaining battery level is low, it is recommended to replace a battery
before it dies.
NOTE
Blinking
<Battery is dead>
[Battery Level] HIGH - - - - - - - - - - - - - - LOW Replace
•
Replace a battery only in non-hazardous locations.
•
Make sure to use a Panasonic alkaline battery (LR03/AAA).
We cannot guarantee the intrinsically safe performance if an
unspecified battery is used. (Applicable battery model is listed
on a label of battery unit.)
•
Make sure to remove dust from O-ring of battery cover and
sealing between the Multi-gas Detector and battery cover, as
dust may be a cause of water ingress into the Multi-gas
Detector.
•
Always remove water and dust before opening the battery
cover. Entry of water or dust inside the Multi-gas Detector may
cause trouble.
•
If an O-ring of battery cover is damaged, be sure to replace
with a new one, as it may lead to water ingress into the
Multi-gas Detector.
- 27 -
4. Error Messages
If an abnormality occurs in the Multi-gas Detector, the error message will be displayed
on the LCD, buzzer will sound intermittently, alarm lamp will flash, and vibrator will start
vibrating intermittently. However the buzzer for sensor error will sound if all of the
mounted sensors are in error.
The buzzer sound will stop by pressing the [ AIR ADJ.] button, and the flashing lamp
and vibration will stop when the power is turned OFF.
Major error messages are listed below. Take appropriate measures in accordance with
the message displayed. When a button or display does not function correctly without
displaying error message, reset battery and turn ON the power again. If it still does not
function properly, contact your New Cosmos representative for repair.
Error Messages
Error Condition Cause
Action
Sensor Error
“E-S” will blink on
the corresponding
gas display.
Sensor Malfunction
If it occurs when the
power is turned ON,
there is a possibility of
presence of gas.
If it occurs when the power is
turned ON, turn OFF the
power and ON again in the
clean air. If it still has a
trouble, ask for repair.
Adjustment Error in
Oxygen Sensor
“E-A” will blink on
the oxygen display.
Oxygen Sensor
Malfunction
Ask for repair.
Battery is Dead
“E-B” will appear.
Battery Voltage
Reduction
This is not a malfunction.
Replace a battery (page 28).
Internal Battery is
Dead
“E-b” will appear.
Battery Voltage
Reduction
Ask for repair.
Detector Error
“E-T” will appear.
Detector Malfunction Ask for repair.
Over Range
“OL” will blink on the
corresponding gas
display.
Exceeding the
Concentration Range
This is not a malfunction.
The gas concentration will be
displayed as the gas
concentration decreases.
(Error in combustible
gas sensor)
▲