38
M1305-0050 Innova CO-170 CO
2
Incubator
User’s Guide
2. If required, you can take a sample of the gas inside the chamber using the CO
2
sample port, and check it using a CO
2
gas analyzer. The CO
2
sample port is
located at the top front of the case, inside the outer door.
If you conduct a sampling, please ensure the following:
Turn off the CO
2
gas by re-programming the setpoint for CO
2
to 0.0% to
prevent CO
2
from being injected into the chamber and giving a false
reading.
A flow rate
≤
0.5 liters/minute is used to take a sample.
The door is kept closed to minimize heat loss.
Remember to reset the CO
2
setpoint to the desired level after sampling.
We recommend that you perform a CO
2
AutoZero prior to sampling (
see Section
3.5
).
3. We also recommend that you AutoZero the CO
2
system at least once every 28
days to ensure that CO
2
level is correct.
Summary of Contents for Innova CO-170
Page 2: ...2 M1305 0050 Innova CO 170 CO2 Incubator User s Guide...
Page 4: ...4 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...
Page 6: ...6 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...
Page 8: ...8 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...
Page 10: ...10 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...
Page 20: ...20 M1305 0050 Innova CO 170 CO2 Incubator User s Guide Figure 4 AutoZero HEPA Filter Fitting...
Page 36: ...36 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...
Page 42: ...42 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...
Page 46: ...46 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...
Page 48: ...48 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...
Page 54: ...54 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...