Rev. 01 on 20/01/2017
20/24
an electromagnetic site. If the force of the magnetic field in the location where the
MAGNETO BIOLIFE device is used exceeds the RF conformity levels mentioned
above, the device should work normally. If you note a functional anomaly,
additional manoeuvres are necessary, such as re-positioning of MAGNETO
BIOLIFE.
b
over the frequency range from 150 kHz to 80 MHz, the forces of the magnetic
field should be less than (V) V/m.
Table 206-Recommended separation distances between portable and mobile RF
telecommunication equipment and the MAGNETO BIOLIFE device
Recommended separation distances between portable and mobile RF
telecommunication equipment and the MAGNETO BIOLIFE device
The MAGNETO BIOLIFE is intended for use in an electromagnetic environment
in which RF radiated disturbance is controlled. The user must avoid
electromagnetic interference by maintaining a minimum distance between
portable and mobile telecommunications equipment and the MAGNETO
BIOLIFE device as recommended above, based on the maximum emission power
of the telecommunications equipment.
Maximum
emission power
of the
transmitters
calculated in
watts (W)
the recommended separation distance in metres based on the
transmitter frequency
√P
√P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters at maximum power not listed above, the recommended distance
“d” in metres “m” can be estimated using the equation applicable to the transmitter
frequency, where “p” is the maximum emission power of the transmitters
calculated in watts (W) according to the manufacturer of the transmitter.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies,
NOTE 2 These guidelines do not apply to all situations, electromagnetic
propagation is influenced by absorption and reflection of structures, objects,
people.