Ca p.6
– M A I N T E N A N C E
6 . 1
– M u l t i - u s e a p p l i c a t o r
Periodically check the connection cable to check for any cracks, a possible cause of
dispersion of magnetic fields; clean the applicator periodically with a damp cloth.
6 . 2
– D e v i c e a n d p o w e r s u p p l y u n i t
Use a damp cloth to clean the power supply unit and the device.
Never use liquids
for any reason, because they are not protected against liquid entering (IP20)
.
6 . 3
– R e p l a c i n g t h e p o w e r s u p p l y u n i t
Check the isolations (casing and cables) on the power supply unit for wear, before
connecting it to the mains. If damaged, even if partially, replace the power supply unit
immediately.
6 . 4
– I m m e d i a t e m a i n t e n a n c e :
Immediate maintenance in New Age Italia or authorised staff if:
the device was subjected to external mechanical stress (e.g. serious fall);
the device was subjected to strong overheating (e.g. if left near intense heat
sources);
if you doubt liquids can penetrate inside;
the power supply unit, the casing and other parts of the device are damaged,
broken or missing;
device functionality appears altered.
For safety purposes, you are advised not to work with accessories (e.g. knobs and
power supply unit) other than those provided as standard equipment.
The frequency for maintenance, functionality testing and correspondence
checking with safety standards EN60601-1 for medical devices should be carried out
annually using a safety tester. The operation lifespan of the device is guaranteed
by the manufacturer only if maintenance is carried out regularly.
NOTE:
you are advised to only have the device tested by New Age Italia. The unit
can be sent directly to the company support lab or handed to the dealer that sold the
device.
Support centre:
New Age Italia srl -
De’ Brozzi, 3 – 48022 Lugo (RA)
Tel:+39-0545.32019 - Telefax: +39.0545.369028 - [email protected]