NEUROTHERM RADIO FREQUENCY LESION GENERATOR
8-7
MODEL NT1100
®
OPERATORS MANUAL
Document 109.00
10September2012 Issue 06
When automatically starting the RF output in
lesion mode with the impedance of probe 2 open
circuit.
Unable to deliver RF power
Impedance open circuit on probe 1.
Check probe and connection of the reference plate
and restart procedure.
Press lesion to go back to procedure screen
3-132
When automatically starting the RF output in
lesion mode with the impedance of probe 3 open
circuit.
Unable to deliver RF power
Impedance open circuit on probe 1.
Check probe and connection of the reference plate
and restart procedure.
Press lesion to go back to procedure screen
3-133
When automatically starting the RF output in pulse
mode with the impedance of probe 1 open circuit.
Unable to deliver RF power
Impedance open circuit on probe 1.
Check probe and connection of the reference plate
and restart procedure.
Press pulse to go back to procedure screen
3-134
When automatically starting the RF output in pulse
mode with the impedance of probe 2 open circuit.
Unable to deliver RF power
Impedance open circuit on probe 2.
Check probe and connection of the reference plate
and restart procedure.
Press pulse to go back to procedure screen
3-135
When automatically starting the RF output in pulse
mode with the impedance of probe 3 open circuit.
Unable to deliver RF power
Impedance open circuit on probe 3.
Check probe and connection of the reference plate
and restart procedure.
Press pulse to go back to procedure screen
3-136
When manually starting the RF output in any
mode with the impedance of probe 1 less than 50
Ohms
Unable to deliver RF power
Impedance is lower than 50 Ohms on probe 1.
Turn off RF power control, check probe and
connection of the reference plate and restart
procedure.
3-137
When manually starting the RF output in any
mode with the impedance of probe 2 less than 50
Ohms
Unable to deliver RF power
Impedance is lower than 50 Ohms on probe 2.
Turn off RF power control, check probe and
connection of the reference plate and restart
procedure.
3-138