NEUROTHERM RADIO FREQUENCY LESION GENERATOR
4-1
MODEL NT1100
®
OPERATORS MANUAL
Document 109.00
10September2012 Issue 06
4
UNPACKING AND ACCEPTANCE TESTING
ON RECEIPT, THE MACHINE SHOULD BE UNPACKED AND INSPECTED FOR ANY
PHYSICAL DAMAGE.
Check that the voltage shown on the rear Serial Number Plate matches the local supply, if not,
contact your local distributor.
Do not attempt to alter the voltage selector on the rear of the
machine.
This is for factory setting only.
Place the NeuroTherm on a flat surface, connect the machine into the mains and switch on using the
ON/OFF switch on the rear of the machine. The following will be observed:
a) The Mains LED will light
b) After a few seconds (during which the computer within the NeuroTherm is going
through a boot sequence) the machine will display the “Self Test” Screen and go
into a self-test routine. As each test is completed the results will be shown on
screen. Upon the completion of all internal tests the Self Test Screen will display
its results for a few seconds.
c) Should a test fail the unit will stop at the fail point with a Warning Message
d) Check Power inlet fuses and Contact NeuroTherm at address in the service section
of this manual
The NeuroTherm will then switch on its “Welcome” Screen
The NeuroTherm is now ready for use. Refer to Section 6 on Getting Started