Neuron-Spectrum-1/V (Technical Manual)
20
moistened with disinfectant (for example, 1% chloramine solution or 3% hydrogen per-
oxide solution). The use of organic solvents and aromatic oils must be avoided. Never
submerge the device or the cables in disinfectant or other liquids.
After testing, the electrodes should be removed from the examined animal immediately.
The disposable needle electrodes should be discarded after use.
5.
Current Repair
5.1. General Requirements
The repair of the system requires special training of technical staff, special equipment
and service software. The manufacturer and its representatives have everything nec-
essary for this. The repair connected with the opening of electronic units is prohibited.
The repair of computer equipment can be performed by companies specialized in
computer equipment servicing.
The current repair of the system includes the repair of some component parts and ca-
bles. If the component parts are connected to the system, the repair is prohibited.
When performing the repair, all units must be switched off.
5.2. EEG Cables and Adapters
The cables are examined externally, and the circuit is checked for short circuit or
break between the screen and the wire and between the wires. In case the cable is
broken or short-circuited, replace it or cut it if the cable length is sufficient.
5.3. Computer Interface Cable (USB Cable)
The computer interface cable (Fig. 8) is examined externally, and the circuit is
checked for short circuit or break. In case the cable is broken, replace it or cut it if the
cable length is sufficient. Please, pay attention for the cable labeling applied along its
entire length. It is prohibited to use the cable with a different labeling.
Fig. 8. Electrical schematic of USB cable.