Neuro-MMG Technical Manua)
12
Requirements to mains:
·
The use of electric mains in which the neutral conductor and protective
ground are combined is strongly prohibited.
·
Before the unit placement, the electrician must check the quality of standard
tripolar sockets and the integrity of the protective ground circuit.
·
In case the MMG unit is connected to system unit of PC or Neuro-IOM elec-
tronic unit plugged to several tripolar sockets, make sure they are grounded
to the same protective ground circuit. Otherwise, there is a danger of several
tens of Ampers leveling current leakage through the unit connecting cables
that leads to the equipment breakdown.
2.3. Unpacking and Delivery Set Check
If the box with the Neuro-MMG was under conditions of the excessive moisture or low
temperature which differs sharply from working conditions, it is necessary to place it in
the room and leave for 24 hours in normal conditions.
Unpack the box and extract the Neuro-MMG. The delivery set should comply with
the packing report list.
Check visually the system components to make sure that there are no visible damag-
es.
2.4. Assembly and Connection to Computer
If you purchase the Neuro-MMG together with the computer, it is delivered with in-
stalled and configured software. If you purchase the Neuro-MMG separately, install
the software using the supplied electronic media (USB flash drive, CD, etc.).
The software must be installed before the first connection of MMG unit to
the computer. Study carefully the corresponding chapter of the Neuro-
IOM.NET or Neuro.MEP.NET user manual before starting to work.
If the distributive is missing, you can download all required software mod-
ules and latest software version from our site:
http://neurosoft.com/en/index/downloads. To get an access to the files, en-
ter login and password that you can inquire in commercial department of
Neurosoft Ltd.