NeurOptics PLR-3000 Instructions For Use Manual Download Page 6

NeurOptics® PLR®-3000 Pupillometer System—Instructions for Use  © 2018 NeurOptics, Inc. 

4

Set Measurement Protocol

From the main screen (Ex. 12), select the 

Settings

 icon   

and then the top right icon   (

Define Protocol

, Ex. 15) to 

navigate to the 

Change Protocol Parameters

 menu (Ex. 16). 

Each parameter listed in the two pages of this menu (Ex. 16) 
can be changed by toggling the values reported using  the 
left “<” and right “>” keys in the directional keyboard.

   Five different experimental protocols can be saved in 

memory and retrieved later when a specific protocol is needed. Press 

Select

 to active a protocol. Use the  

left “<” and right “>” keys in the directional keyboard to change the protocol number. Use the list of parameters 
below to change the settings of the selected protocol.

   Use this setting to change the type of the experiment: 

Positive Pulse Stimulus: 

standard pupil light reflex, a bright pulse over a dimmer background triggers a pupil constriction. 

Negative Pulse Stimulus: 

dim pulse over a brighter background triggers a dark reflex and pupil dilation. 

Static Stimulus:

 no light stimulation and no pupil reflex, just a continuous and static light background.  

Extended Mode:

 pupil is recorded at a lower frequency for an extended period of time (maximum duration 10 minutes). 

   

Pulse Intensity.

 Use this setting to change the intensity of the light stimulus. Unit of light is radiometric and provided 

in micro Watts, uW. Five different intensities are provided: 0uW, 1uW, 10uW, 50uW, 121uW, and 180uW. 

   

Background Intensity. 

Use this setting to change the background light intensity. Note that in case of a Positive Pulse 

Stimulus protocol, Background Intensity must be less than Pulse Intensity whereas, for a Negative Pulse Stimulus, 
protocol it is the opposite -- Background must be higher than the Pulse Intensity.

   

Measurement Duration.

 This setting is for setting the duration of the measurement, a minimum of 3 seconds up  

to a maximum of 41 seconds.

   

Pulse Duration.

 This is the duration of the light pulse stimulation, a minimum of 0.03 sec for the Positive Pulse  

and 1 sec for the Negative Pulse.

   

Pulse Onset.

 This is the start of the Pulse stimulation. 

In case there is a conflict in the settings, for example the Background Intensity is set brighter than the Pulse Intensity  
in a Positive Pulse Stimulus protocol, then the corresponding icons are shown in red and the Define Protocol menu 
cannot be exited until the conflict is resolved. 

Measure Pupils

The PLR-3000 should not be used without the eye cup positioned correctly (Ex. 17). It is  
very important that the eye cup be correctly fitted. A snug fit helps reduce the possibility  
of stray light entering the eye while the scans are taking place. The eye cup has a tab in  
the rim, which fits into the indentation in the lens shield of the Pupillometer. Position the  
tab in the eye cup rim into the indent in the lens shield and press into place. The tabs on  
either side of the lens shield should also snap into the holes on either side of the eye cup.

Position the PLR-3000 with the eye cup at a right angle  
to the patient’s axis of vision, minimizing any tilting of the  
device (Ex. 18).

Ex. 18

Ex. 17

Ex. 15

Ex. 16

Summary of Contents for PLR-3000

Page 1: ...with a color touchscreen LCD screen for data display A keypad completes the user interface and enables manual entry of individual patient identification ID Indications for Use The PLR 3000 Pupillomet...

Page 2: ...Medical Products These limits provide reasonable protection against electromagnetic interference when operated in the intended use environments e g hospitals research laboratories Magnetic Resonance I...

Page 3: ...Set Measurement Protocol 4 Measure Pupils 4 Browse Records 5 Download Records 6 Power Down 6 Troubleshooting 6 Cleaning and Maintenance 7 Ordering Information 8 Customer Service 9 Appendix A Pupillary...

Page 4: ...ometer Power Supply to the PLR 3000 Charging Station and plug into a power outlet The green light at the base of the Charging Station will indicate power has been established Ex 2 Place the PLR 3000 i...

Page 5: ...is heard and a blue light on the device flashes Position the Barcode Scanner next to the PLR 3000 On the PLR 3000 select Barcode Scanner The PLR 3000 will display Connecting on the touchscreen Ex 9 O...

Page 6: ...nd 180uW Background Intensity Use this setting to change the background light intensity Note that in case of a Positive Pulse Stimulus protocol Background Intensity must be less than Pulse Intensity w...

Page 7: ...e latency and the blue line the T75 Latency and T75 are two of the variables calculated by the analysis and they are explained in Appendix A If a variable could not be calculated for example because o...

Page 8: ...File or In Windows 7 click on the Start button and type fsquirt in the box that says Search programs and files In Windows 8 and 8 1 open the Start screen and type fsquirt In Windows 10 right click Sta...

Page 9: ...prior to wiping the pupillometer and charging station and do not use an oversaturated cloth Wipe all exposed surfaces Follow the cleaner s manufacturer instructions as to the time required to leave t...

Page 10: ...illometer and the gold connector pins located in the base of the PLR 3000 Charging Station the cleaning process should be followed by a second cleaning using 70 IPA solution to ensure that all residue...

Page 11: ...ll Free North America 866 99 PUPIL 866 997 8745 p 949 250 9792 International 1 949 250 9792 or email Support neuroptics com Returned Goods Policy Products must be returned in unopened packages with ma...

Page 12: ...me of onset of constriction following initiation of the light stimulus ACV Constriction Velocity Average velocity of how the pupil diameter is constricting measured in millimeters per second MCV Maxim...

Page 13: ...battery operation Power Supply Input 100 240 VAC 8 Output 6V 2 8 Amps Battery 3 7V 3350 mAh hour Li Ion Cell Operating Environment Temperature Range 18 C 65 F to 30 C 86 F Relative Humidity 20 to 70...

Page 14: ...18101 Von Karman Avenue Suite 1940 Irvine CA 92612 USA p 949 250 9792 Toll Free North America 866 99 PUPIL info NeurOptics com NeurOptics com PLR3_IFU Rev D HOMA 9YGRHF...

Reviews: