Cleaning Considerations: Gold Connector Pins and Blades

In instances where there is concern of exposure to highly resistant bacteria or viruses (ie: Clostridium difficile, or “C. 
diff”), we understand that hospital protocols may require use of cleaning solutions containing sodium hypochlorite 
(bleach) when cleaning equipment. Please be aware solutions containing sodium hypochlorite (bleach) will corrode 
the gold connector blades located on the bottom of the PLR-3000 Pupillometer handle (Figure 26), and the gold 
connector pins located in the base of the PLR-3000 Charging Station (Figure 27.) 
 
 

 
 
 
 
 
 
 
 
 
 
 

 

• 

DO NOT

 

use products containing sodium hypochlorite (bleach) to clean the gold connector blades located    

 

  on the bottom of the PLR-3000 Pupillometer handle, and the gold connector pins located in the base of the  

 

  PLR-3000 Charging Station

If products containing sodium hypochlorite (bleach) are used to clean the gold connector blades located on the 
bottom of the PLR-3000 Pupillometer and the gold connector pins located in the base of the PLR-3000 Charging 
Station, the cleaning process should be followed by a second cleaning using 70% IPA solution to ensure that all 
residue is completely removed from the device in order to minimize damage to the gold connector pins and blades. 

Cleaning Considerations: Pupillometer Liquid Crystal Display (LCD)

For best protection of the liquid crystal display (LCD), use a clean, soft, lint-free cloth and 70% IPA cleaning solution to 
clean the pupillometer optics.

In instances where there is concern of exposure to highly resistant bacteria or viruses (ie: Clostridium difficile, or “C. 
diff”), we understand that hospital protocols may require use of cleaning solutions containing sodium hypochlorite 
(bleach) when cleaning equipment. If products containing sodium hypochlorite (bleach) are used to clean the LCD of 
the PLR-3000 Pupillometer, the cleaning process should be followed by a second cleaning solution with a 70% IPA 
solution to ensure that all bleach residue is completely removed from the LCD using a clean, soft, lint-free cloth.

Figure 26- Gold Connector Blades

Figure 27- Gold Connector Pins

Gold

Connector 

Blades

Gold

Connector 

Pins

PLR-3000 Pupillometer 

PLR-3000 Pupillometer 

Summary of Contents for PLR-3000

Page 1: ...with a color touchscreen LCD screen for data display A keypad completes the user interface and enables manual entry of individual patient identification ID Indications for Use The PLR 3000 Pupillomet...

Page 2: ...Medical Products These limits provide reasonable protection against electromagnetic interference when operated in the intended use environments e g hospitals research laboratories Magnetic Resonance I...

Page 3: ...Set Measurement Protocol 4 Measure Pupils 4 Browse Records 5 Download Records 6 Power Down 6 Troubleshooting 6 Cleaning and Maintenance 7 Ordering Information 8 Customer Service 9 Appendix A Pupillary...

Page 4: ...ometer Power Supply to the PLR 3000 Charging Station and plug into a power outlet The green light at the base of the Charging Station will indicate power has been established Ex 2 Place the PLR 3000 i...

Page 5: ...is heard and a blue light on the device flashes Position the Barcode Scanner next to the PLR 3000 On the PLR 3000 select Barcode Scanner The PLR 3000 will display Connecting on the touchscreen Ex 9 O...

Page 6: ...nd 180uW Background Intensity Use this setting to change the background light intensity Note that in case of a Positive Pulse Stimulus protocol Background Intensity must be less than Pulse Intensity w...

Page 7: ...e latency and the blue line the T75 Latency and T75 are two of the variables calculated by the analysis and they are explained in Appendix A If a variable could not be calculated for example because o...

Page 8: ...File or In Windows 7 click on the Start button and type fsquirt in the box that says Search programs and files In Windows 8 and 8 1 open the Start screen and type fsquirt In Windows 10 right click Sta...

Page 9: ...prior to wiping the pupillometer and charging station and do not use an oversaturated cloth Wipe all exposed surfaces Follow the cleaner s manufacturer instructions as to the time required to leave t...

Page 10: ...illometer and the gold connector pins located in the base of the PLR 3000 Charging Station the cleaning process should be followed by a second cleaning using 70 IPA solution to ensure that all residue...

Page 11: ...ll Free North America 866 99 PUPIL 866 997 8745 p 949 250 9792 International 1 949 250 9792 or email Support neuroptics com Returned Goods Policy Products must be returned in unopened packages with ma...

Page 12: ...me of onset of constriction following initiation of the light stimulus ACV Constriction Velocity Average velocity of how the pupil diameter is constricting measured in millimeters per second MCV Maxim...

Page 13: ...battery operation Power Supply Input 100 240 VAC 8 Output 6V 2 8 Amps Battery 3 7V 3350 mAh hour Li Ion Cell Operating Environment Temperature Range 18 C 65 F to 30 C 86 F Relative Humidity 20 to 70...

Page 14: ...18101 Von Karman Avenue Suite 1940 Irvine CA 92612 USA p 949 250 9792 Toll Free North America 866 99 PUPIL info NeurOptics com NeurOptics com PLR3_IFU Rev D HOMA 9YGRHF...

Reviews: