INSTALLING NEW BATTERIES

To  access  the  batteries,  push 
down  and  slide  open  the  access 
door  on  the  top  of  the  instrument. 
Replace  the  batteries  according 
to  the  diagram  on  the  label  on  the 
back  of  the  instrument.  Use  only  4 
rechargeable AA-size NiMH batteries 
(Neogen  part  #9610).  Old  batteries 
may be disposed of as nonhazardous 
waste. 

NOTE:

  If  the  batteries  are 

removed for more than 1 minute, the 
time and date must be reset, but the 
data readings will be retained.

SLEEP MODE

When  not  being  actively  used 
for  more  than  five  minutes,  the 
AccuPoint  2  instrument  will  enter 
“sleep”  mode  and  dim  the  color  display  to  conserve  power.  Depressing  the  toggle 
button causes the instrument to return to active mode. 

COMMUNICATING WITH THE DATA MANAGER SOFTWARE

The  AccuPoint  2  instrument  can  communicate  with  a  computer 
through the Data Manager software to upload the Test Plan and User 
information to the instrument or download the testing results from the 
instrument. All  communication  with  the  instrument  is  accomplished 
through the supplied USB cable. The USB connection along with the 

AC adapter port are located behind the access door on the side of the AccuPoint 2 
instrument. When the cable is plugged in and a connection is established, the graphic 
above will appear on the instrument’s display. 

DOWNLOADING TEST RESULTS TO THE DATA MANAGER SOFTWARE

The AccuPoint 2 instrument can store up to 4,000 data records. The instrument will 
display the number of stored data records on the User ID screen at power-up. Once the 
limit has been reached, the oldest record will be discarded as each new record is added. 
To download data to a computer:
1.   Open the AccuPoint Data Manager program on the computer (if necessary, see 

the software manual for program installation instructions). 

2.  Turn on the instrument and plug in the USB cable. If the USB cable is not connected 

at the other end to your computer connect it now.

3.   In the File menu, select Load Data. A dialog box will be displayed.

Battery access door. Push 
down and slide toward back 
of the instrument to access 
batteries.

04:22P    

01/10/2010

Jose

001

Line 1

Conveyor Belt  1

135

0011

Pass

Marginal

Fail

Liquid sampling mode 

Display contrast 

 

 

Reading Sampler 

 

USB connection 

established

Sampler not detected

Sampler door open

Date

User

Time

Site 

Number

Site 

Group

Site 

Name

Test 

Result

Record 

Count

The correct battery 
orientation is shown 
on the back of the 
instrument (below)

AP2 Users Guide1a.indd   11

10/20/2009   3:47:45 PM

Summary of Contents for AccuPoint 2 ATP

Page 1: ...t 2 system 10 Recharging the batteries 10 Installing new batteries 11 Sleep mode 11 Communicating with the Data Manager software 11 Downloading test results to the Data Manager software 11 Setting thresholds in the instrument 12 Cleaning the instrument 12 Troubleshooting 13 Regulatory information 14 Replacement parts accessories 15 Warranty 15 Customer service contact information 15 Neogen Corpora...

Page 2: ...2 system utilizes ATP bioluminescence to determine the cleanliness of test samples Adenosine triphosphate ATP is a chemical compound found in all living cells including bacteria food debris yeast and mold Bioluminescence is a chemical reaction that produces light ATP bioluminescence occurs when ATP from a sample comes into contact with luciferase an enzyme found in fireflies and luciferin a substr...

Page 3: ...ccuPoint 2 system Refrigerate the sampler cartridges at 2 8 C 35 46 F Within this temperature range cartridges may be stored until their expiry Sampler cartridges must be warmed to room temperature prior to use After removing sampler cartridges from the bag be sure to reseal and refrigerate the bag at 2 8 C Do not touch the tip of the sampler or let the tip touch any other surface prior to testing...

Page 4: ...esired language to the back of the instrument Before the first use of the system the batteries should be fully charged Connect the A C power cord to a power supply and allow the batteries to charge for approximately 90 minutes before use It is not necessary to recharge the instrument after each use Switch on the instrument using the red power button located on the front of the instrument The instr...

Page 5: ...on where the test is performed Test Result The relative light unit RLU score for this test This is the result that will be used to determine whether the test indicates a pass marginal or fail Compared to the thresholds that were established in the Data Manager software for this site the results are Pass The surface or liquid sample is considered clean Marginal The sample is considered between the ...

Page 6: ...he appropriate site before taking each test The result along with the time date user site name and site group can then be downloaded to the software for tracking and analysis SELECTING THE TESTING SITE If a test plan has been uploaded from the Data Manager software toggle to the appropriate site through the Site Number Site Group or Site Name fields The Site Name field has an option for selecting ...

Page 7: ... the left will decrease the contrast RECORD COUNT AND TEST RECALL The instrument keeps track of the number of tests that have been taken The number of records currently stored in the instrument s memory is displayed at the User ID screen when the instrument is powered on The number of tests that have been taken since the instrument was last powered on is displayed in the bottom right corner of the...

Page 8: ...o ensure accurate test results do not touch the tip of the sampler or let the tip touch any other surface prior to testing 4 Surface sampling Sample a surface of approximately 10 cm by 10 cm 4 in by 4 in Sample by starting in one corner of the square and forming a continuous line going back and forth across the square Repeat the procedure with slightly less pressure from a starting point 90 from t...

Page 9: ...l begin immediately An icon like the one below will appear on the screen along with a progress indicator bar 8 The reading is displayed along with a symbol indicating the result The symbols are identified below 9 Press the eject button and remove the sampler from the chamber Samplers and their cartridges are non toxic and may be disposed of without any special considerations Pass Marginal Fail AP2...

Page 10: ...s should only be recharged when the battery charge icon appears near depletion The AccuPoint 2 instrument can only be recharged through the supplied A C power adapter Full charge takes approximately 90 minutes After inserting the A C power cord the instrument will display a green battery with the appropriate charge state This icon will change and fill in completely as the batteries become fully ch...

Page 11: ...ted behind the access door on the side of the AccuPoint 2 instrument When the cable is plugged in and a connection is established the graphic above will appear on the instrument s display DOWNLOADING TEST RESULTS TO THE DATA MANAGER SOFTWARE The AccuPoint 2 instrument can store up to 4 000 data records The instrument will display the number of stored data records on the User ID screen at power up ...

Page 12: ...ete uploading it to the instrument through the Download Thresholds command will establish the correct thresholds for each site If a Test Plan will not be used at a facility and to simply operate the system in a pass fail mode it will still be necessary to upload a simple test plan to not use the default thresholds Simply establish a single test site in the Test Plan Setup routine with its appropri...

Page 13: ...nt is too cold Allow the instrument to warm to ambient temperature The instrument is too hot Allow the instrument to cool to ambient temperature The power was turned on Press the Eject button and while there is still a sampler remove the sampler in the chamber The battery charge is too low Recharge the batteries The instrument has lost its Contact Neogen support serial number AP2 Users Guide1a ind...

Page 14: ...emperature minus 20 60 C minus 4 140 F Relative humidity 20 80 non condensing Altitude 0 12 000 m AccuPoint 2 instrument The power supply is designed for indoor use only AC line voltage frequency 110 240 VAC 47 63 Hz Maximum input current 1 amp Installation category This instrument is able to withstand transient overvoltage according to Installation Category II as defined in IEC 1010 1 Pollution d...

Page 15: ...anship Buyer assumes all risk and liability resulting from the use of this product There is no warranty of merchantability of this product or of the fitness of the product for any purpose Neogen shall not be liable for any damages including special or consequential damage or expense arising directly or indirectly from the use of this product CUSTOMER SERVICE CONTACT INFORMATION Customer assistance...

Page 16: ..._____________________________ _______________________________________________________________ _______________________________________________________________ _______________________________________________________________ _______________________________________________________________ _______________________________________________________________ __________________________________________________...

Reviews: