Installation
Service Manual
12
Oxinet III
Installation
The Oxinet III system should only be installed by trained Nellcor personnel or
authorized representatives. The installation information provided in this manual
can be used to reconnect the Oxinet III system after moving it.
Wired Configuration
The following Nellcor oximeters are supported in the Oxinet III system:
• N-395
• N-550
• N-595
• N-560
• N-600
The Oximeter data output port must be configured as follows:
• communication protocol — ASCII
• baud rate — 9600
Refer to the Oximeter’service manual for details on Oximeter settings.
Connect the Oximeter to the Communication Server
1
Plug a DB15M/RJ-45 Cat5 Adapter connector into the data port of each Oximeter
(patient room/bed) as shown in Figure 3.
2
Connect a AC power cord to the power input jack on the back of each Oximeter
and plug the other end into AC power outlet.
3
Connect a Cat5 Patch cable from the DB15M/RJ-45 connector of each Oximeter
to an available serial port on the back of the Communication Server and take note
of the room sequence versus the communication serial port number (Figure 3).
Connect the Communication Server to the Router/Switch
1
Connect the Cat5 Patch cable from the 10/100 Base-T network port on the
Communication Server to any available numbered RJ-45 LAN port on the Router/
Switch as shown in Figure 3 (
DO NOT use the port labeled “WAN”
).
2
Connect a AC power cord to the power input jack on the back of the
Communication Server and plug the other end into one of the battery backed-up
AC power outlets of the UPS.
3
Connect the DC barrel connector end of the DC Adapter Power cable to the round
DC jack on the back of the Router/Switch and plug the AC end into one of the
battery backed-up AC power outlets of the UPS (Uninterruptible Power Supply).
Summary of Contents for Oxinet III
Page 10: ...Introduction Service Manual 4 Oxinet III...
Page 20: ...Installation Service Manual 14 Oxinet III Figure 3 Wired Configuration System Connection...
Page 24: ...Installation Service Manual 18 Oxinet III Figure 4 Wireless Configuration System Connection...
Page 44: ...Administrative Functions Service Manual 38 Oxinet III...
Page 48: ...Maintenance Service Manual 42 Oxinet III...
Page 58: ...Specifications Service Manual 52 Oxinet III...
Page 61: ......