Troubleshooting
48
Operational Performance
Table 9 lists symptoms of problems relating to operational performance (no
error codes displayed) and recommended actions. If the action requires
replacement of a PCB or module, refer to
An alarm condition
exists but no alarm
(audible or visual) is
indicated.
Replace the Front Panel PCB. Refer to
Table 8: Display/Alarms Problems (Continued)
Symptom
Recommended Action
Table 9: Operational Performance Problems
Symptom
Recommended Action
The pulse amplitude
indicator seems to
indicate a pulse, but the
digital displays show
zeroes.
The sensor may be damaged; replace it.
Replace the Front Panel PCB. Refer to
SpO
2
or pulse values
change rapidly; Pulse
Amplitude indicator is
erratic.
The sensor may be damp or may have been reused too
many times. Replace it.
•
An electrosurgical unit (ESU) may be
interfering with performance:
•
Move the N-550 and its cables and sensors as
far from the ESU as possible.
•
Plug the N-550 power supply and the ESU
into different AC circuits.
•
Move the ESU ground pad as close to the
surgical site as possible and as far away from
the sensor as possible.
•
Verify the performance with the procedures
detailed in
•
Replace the Front Panel PCB. Refer to
Summary of Contents for OxiMax N-550
Page 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Page 10: ...Blank Page ...
Page 38: ...Blank Page ...
Page 64: ...Blank Page ...
Page 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Page 82: ...Blank Page ...
Page 86: ...Spare Parts 76 Figure 40 Exploded View ...
Page 90: ...Blank Page ...
Page 108: ...Blank Page ...
Page 120: ...Blank Page ...
Page 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Page 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Page 136: ...Blank Page ...
Page 137: ... ...