P e r f o r m a n c e V e r i f i c a t i o n
Introduction
This section discusses the tests used to verify performance following repairs
or during routine maintenance. All tests can be performed without removing
the N-550 cover. All tests except the battery charge and battery performance
tests must be performed as the last operation before the N-550 is returned to
the user.
If the N-550 fails to perform as specified in any test, repairs must be made to
correct the problem before the N-550 is returned to the user.
Equipment Needed
Table 1: Equipment Needed
Equipment
Description
Digital multimeter (DMM)
Fluke Model 87 or equivalent
Durasensor
®
oxygen transducer
(sensor)
DS-100A
O
XI
M
AX
oxygen transducer
MAX-A
Pulse oximeter tester
SRC-MAX (must meet current EN
61326-1, FCC Part 15, and UL 1010-1,
or IEC 61010-1, or CSA 22.2 No.
1010-1 specifications)
Safety analyzer
Must meet current AAMI ESI/1993 &
IEC 60601-1/1998 specifications
Pulse oximetry cable
DOC-10
Data interface cable
EIA-232 cable (optional)
Stopwatch
Manual or electronic
Summary of Contents for OxiMax N-550
Page 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Page 10: ...Blank Page ...
Page 38: ...Blank Page ...
Page 64: ...Blank Page ...
Page 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Page 82: ...Blank Page ...
Page 86: ...Spare Parts 76 Figure 40 Exploded View ...
Page 90: ...Blank Page ...
Page 108: ...Blank Page ...
Page 120: ...Blank Page ...
Page 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Page 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Page 136: ...Blank Page ...
Page 137: ... ...