Appendix
A-4
A4.2 Patient Data and Operating Status
Time Tag
Time Tag
%SpO2
PR (bpm)
PA
Status
123456789
100
120
220
Time Tag does not represent a real-time clock. The number beneath the Time
Tag heading represents time, in seconds, since the unit was initialized at the
factory. This number will increase in size throughout the life of the monitor.
Patient Data
NPB-190
Version 1.0.0
CRC XXXX
SpO2 Limit: 30-100%
PR Limit: 100-180 bpm
Time Tag
%SpO2
PR (bpm)
PA
Status
123456795
100
190*
220
PH
Patient data are highlighted in the display above. Parameter values are displayed
directly beneath the heading for each parameter. In this example, the %SpO
2
is
100, and the pulse rate PR is 190 beats per minute. The “*” next to the 190
indicates that 190 beats per minute is outside of the alarm limits, indicated in the
top row, for pulse rate. If no data for a parameter is available three dashes (- - -)
will be displayed in the printout.
PA is an indication of pulse amplitude. The number can range from 0 to 254 and
will typically range around 45. There are no alarm parameters for this value. It
can be used for trending information. It is an indication of a change in pulse
volume, pulse strength, or circulation.
NPB-190
Version 1.0.0
CRC XXXX
SpO2 Limit: 30-100%
PR Limit: 100-180 bpm
Time Tag
%SpO2
PR (bpm)
PA
Status
123456795
100
190*
220
PH
The Status column indicates alarm conditions and operating status of the NPB-
190. A Pulse High alarm is indicated by the PH in this example. The status
column can have as many as four codes displayed in one line of data. A complete
listing of the status codes is listed in Table A-2.
Table A-2: Status Codes
Code
Meaning
LB
L
ow
B
attery
AS
A
larm
S
ilence
AO
A
larm
O
ff
SD
S
ensor
D
isconnect
PS
P
ulse
S
earch
LP
L
oss of
P
ulse
SH
S
at
H
igh Limit Alarm
SL
S
at
L
ow Limit Alarm
PH
P
ulse rate
H
igh Limit Alarm
PL
P
ulse rate
L
ow Limit Alarm
Note: A Sensor Disconnect will also cause three dashes (- - -) to be displayed
in the patient data section of the printout.