![Nellcor NPB-190 Service Manual Download Page 55](http://html.mh-extra.com/html/nellcor/npb-190/npb-190_service-manual_718065055.webp)
9-1
SECTION 9: SPECIFICATIONS
9.1 General
9.2 Electrical
9.3 Physical Characteristics
9.4 Environmental
9.5 Alarms
9.6 Factory Default Settings
9.7 Performance
9.1
GENERAL
Designed to meet safety requirements of:
UL 2601-1 CSA-C22.2 No. 601-1-M90, IEC 601-1 (Class I, type BF)
EMC per EN60601-1-2
9.2
ELECTRICAL
Protection Class
Class I: per IEC 601-1, clause 2.2.4
Degree of Protection
Type BF: per IEC 601-1, clause 2.2.25
Mode of Operation
Continuous
Battery
Type
Rechargeable, sealed lead-acid, internal
Operating time
12 hours minimum on new, fully charged
battery
Recharge period
14 hours for full charge
Fuses
2 each
5 X 20 mm
Slow Blow. 0.5 amp, 250 volts
AC Power
Selectable by switch
100-120 VAC 50/60 Hz or
200-240 VAC 50/60 Hz