Intended Use

The MD215MG grayscale display is intended to be used for displaying and viewing of digital images for diagnosis, including 
digital mammography, by trained physicians.

Caution: 

The MD215MG grayscale display cannot be used for a life-support system.

 

This device is designed as component of a fi nal system which is in compliance to IEC60601-1-1 requirements.

Safety Precautions

This device is designed for exclusive interconnection with IEC60601-1 certifi ed equipment outside of patient vicinity.

•  Equipment connected to digital interfaces must comply with the respective IEC standards (e.g. IEC 60950 for data processing 

equipment and IEC 60601-1 for medical equipment). 

•  This device complies with EN60601-1-2. To minimize the interference from other equipment, a minimum 0.5 m distance shall 

be kept form other potential electromagnetic sources, such as a Cell Phone. 

•  To reduce the risk of electric shock, make sure that the power cord is unplugged from wall socket. To fully disengage the 

power to the unit, please disconnect the power cord from the AC outlet. Do not open the back cover or AC adapter. No user 
serviceable parts inside. Refer servicing to qualifi ed service personnel. The AC outlet shall be readily available and accessible.

North America Customers

Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle, marked “Hospital Only” 
or “Hospital Grade”. That signal input port need to be connected properly and that any unused signal input port shall not be 
accessible for the patient to contact in the patient area after the LCD is integrated into a medical system.
Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner.

European Customers

Unpacking, Installation and calibration of this display should only be done by authorized and trained personnel. Any installation 
done by a non-authorized person is done under his own risk and we accept no responsibility in any device malfunctioning. To 
secure full functionality of this display, only NEC approved display controller models for connection with the fi nal system must be 
used.

Medical Imaging

MD215MG is designed for 2048 x 2560 medical imaging.

Summary of Contents for MultiSync MD215MG-S5

Page 1: ...MD215MG...

Page 2: ...and or registered trademarks of Microsoft Corporation in the United States and or other countries VESA is a trademark of the Video Electronics Standard Association All other products and company name...

Page 3: ...English...

Page 4: ...erts the user that important literature concerning the operation and maintenance of this unit has been included Therefore it should be read carefully in order to avoid any problems Protective earth te...

Page 5: ...e equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the...

Page 6: ...DE REFER SERVICING TO QUALIFIED SERVICE PERSONNEL CAUTION Power Cord Important Information CAUTION Please use the power cord provided with this display in accordance with the table below If a power co...

Page 7: ...wer to the unit please disconnect the power cord from the AC outlet Do not open the back cover or AC adapter No user serviceable parts inside Refer servicing to qualified service personnel The AC outl...

Page 8: ...service personnel under the following conditions If the monitor is used in this condition the monitor may cause fall fire and electric shock If the monitor stand has been cracked or peeled If the moni...

Page 9: ...directly in front of you to minimize turning your head Avoid displaying fixed patterns on the monitor for long periods of time to avoid image persistence after image effects Get regular eye checkups...

Page 10: ...le 1 AC adapter MW122 Power Cord 2 Quick Reference Guide USB Cable Remember to save your original box and packing material to transport or ship the monitor 1 Dual link DVI cable 2 Type and number of p...

Page 11: ...gnal cable to the connector of the display controller in your system Figure A Tighten all screws Figure A 3 Remove the cable cover Figure B To remove the cable cover unlock the slide latch 1 on the ca...

Page 12: ...lled the cables NOTE Handle with care when raising lowering rotating or tilting the monitor screen 8 Connect the AC adapter with the proper AC cord then plug the AC cord to the power outlet NOTE Pleas...

Page 13: ...process 5 plus Button To open the OSD On Screen Display 1st click Display information 2nd click Present operating state 3rd click OSD off otherwise OSD disappears in 15 sec 6 minus Button Same as plus...

Page 14: ...l Power saved Approx 90W w supplied AC adapter Less than 10W w supplied AC adapter Dimensions weight Net W x H x D Packed 390 x 487 8 584 x 227 6 mm W x H x D Portrait approx 11 1 kg 475 x 705 x 345 m...

Page 15: ...e of a customized driver for the display controller Make sure that a customized driver is available for your specific display controller model For the availability of a packed pixel driver please cont...

Page 16: ...Click the plus button or minus button on the back side of the monitor once to display the INFORMATION OSD A display which is similar to this picture opens on the screen 2 Hold down the plus button and...

Page 17: ...rator Manual 1 Click the plus button or minus button on the back side of the monitor twice to display the PRESENT STATE OSD A display like this picture opens on the screen 2 Hold down the plus button...

Page 18: ...m The tilt and swivel stand as shipped with the display is interchangeable with commercially available wall mounts or multi display mounting arms that Come with a mount plate with 4 slotted screw hole...

Page 19: ...ay is equipped with the security slot supported by MicroSaver Security Lock System by Kensington Technology Group The MicroSaver locks are available at PC supply stores Note Kensington is the register...

Page 20: ...s http www nec display solutions com in Europe http www nec display com in Japan or http www necdisplay com in USA Energy Saving This monitor features an advanced energy saving capability Mode Power c...

Page 21: ...8 0023 Japan NEC Display Solutions of America 500 Park Blvd Suite 1100 Itasca Illinois 60143 USA Phone 630 467 3000 Fax 630 467 3010 NEC Display Solutions Europe GmbH Landshuter Allee 12 14 D 80637 Mu...

Reviews: