48
49
7|
Technial data
Technial data
|7
7|
7.2. Aerosol delivery rate and aerosol spectrum
Aerosol delivery rate
When using the SaHoMa®-II control unit in connection with the SaHoMa®-II mesh nebulization unit, the
average
aerosol delivery rate
is 0.8 ml isotonic saline solution (NaCl) 0.9 % per minute. The aerosol delivery
rate of a SaHoMa®-II mesh nebulization unit is within tolerance if it is 0.5 to 3 ml of isotonic saline solution
(NaCl) 0.9 % per minute.
Aerosol spectrum
When using the SaHoMa®-II control unit in combination with the SaHoMa®-II mesh nebulization unit (test
medium: isotonic saline solution (NaCl) 0.9 %), the
aerosol spectrum
was determined as follows:
Determination of the aerosol spectrum using a laser diffractometer
(HELOS VARIO KF)
The average particle size is
5.14 µm MMD
with a GSD of 1.91.
48.28 % of the particles are ≤ 5 µm.
Technical data
7.1. Technical data of the SaHoMa®-II inhalation set for horses
Technical data of the SaHoMa®-II control unit
Dimensions (L x W x H):
112 x 82 x 66 mm
Weight:
208 g
Power supply types:
Battery operation
Type of battery:
Lithium polymer battery / 11,1 V DC / 1200 mAh
Power cord:
100 - 240 V AC; 50/60 Hz / 12 V; 1,5 A
Technical data of the SaHoMa®-II mesh nebulization unit
Dimensions (L x W x H):
51 x 33 x 77 mm
Weight:
31 g
Maximum filling quantity:
17 ml
For further information about performance and particle size please refer to
chapter 7.2 „Aerosol delivery rate
and aerosol spectrum“.
Technical data of the SaHoMa®-II inhalation set for horses in working conditions
Dimensions (L x W x H):
260 x 250 x 322 mm
Weight:
1,45 kg