FAQ & self help
FAQ & self help
|6
Error
Possible Causes / Elimination
The inhalation times are
noticeably longer
and there is
no or only
a small aerosol
output.
Possible Causes:
• Dirty or clogged membrane of the SaHoMa®-II mesh
nebulization unit
• Mechanically damaged membrane of the SaHoMa®-II mesh
nebulization unit (e. g. in case of improper handling)
• The end of the lifespan of the SaHoMa®-II mesh nebulization unit
has been reached or exceeded
• The room or ambient temperature is below 10° C
Elimination:
• Cleaning the SaHoMa®-II mesh nebulization unit
(chapter 4.2)
• Replacement of the SaHoMa®-II mesh nebulization unit
(chapter 9)
• Please wait approx. 30 minutes until the system has adjusted to
room temperature and only fill and nebulize inhalation solutions
that are at room temperature.
• Connect a SaHoMa®-II mesh nebulization unit filled with isotonic
saline solution (NaCl 0.9 %), allow the device to nebulize for
10 seconds and then switch it off again manually and switch it on
again in the desired program
The SaHoMa®-II control unit
switches off
unintentionally
during the inhalation process despite
a remaining residual quantity.
Possible Causes:
• Air bubbles on the membrane of the SaHoMa®-II mesh
nebulization unit
• Use of a foaming medication suggesting a filling in form
of foam residues
Elimination:
• Light tapping of the SaHoMa®-II
• mesh nebulization unit to loosen the air bubbles;
if necessary restart the SaHoMa®-II control unit
• Proper dilution of foaming medications according to the list
of medications; nebulization in program 2 under constant obser-
vation with manually switching off the SaHoMa®-II control unit.
• Connect a SaHoMa®-II mesh nebulization unit filled with isotonic
saline solution (NaCl 0.9 %), allow the device to nebulize for 10
seconds and then switch it off again manually and switch it on
again in the desired program
Error
Possible Causes / Elimination
Charging
the
SaHoMa®-II control unit
is
not possible.
Possible Causes:
• Defective SaHoMa®-II power cord or defective plug of the
SaHoMa®-II power cord
• No voltage at the socket used
• Defective charging socket
Elimination:
• Use of another socket
• Check the plug connection at the SaHoMa®-II control unit
• Otherwise send in for inspection/repair
The SaHoMa®-II control unit
cannot be switched off
and
does not respond to pressing
the button during operation.
Possible Causes:
• The button on the SaHoMa®-II control unit has been pressed
repeatedly within a short period of time
• The SaHoMa®-II control unit no longer responds to a button press
due to an unusually high workload
Elimination:
• Put the SaHoMa®-II control unit aside for approx. 10 minutes until
it automatically switches itself off and performs a reset
Warning
During this time, the SaHoMa®-II control unit must not be
connected for charging!
Service
In all other cases or in case of uncertainty, please contact our service staff who will be happy to help you:
+49 (0) 6022 - 610 62 -0.
Please note that in the case of a repair, you will have to complete the repair form and send it to us with all the