7
3
.0 P
reparing the Device
Check component functionality
Before use, check that the device is working
properly. To do this, the following procedure should
be carried out:
•
Connect the power supply and check that the
green LED is on the docking station
•
Take out the PLASMA IQ from the docking
station, press the LOW button and check if the
yellow LED turns on
•
Take out the PLASMA IQ from the docking
station, press the HIGH button and check that
the yellow LED lights up
On the device, there
are two buttons
used to provide
two different output
voltages, marked
with inscriptions
“H” high and “L” low.
Figure 1: Device PLASMA IQ
Preparing the device and equipment for work
Before starting work with the device, an active electrode should be inserted. In order to install
the active electrode, ensure that the device is off and follow the steps below and the drawing:
1
Turn the spindle to the left;
2
Insert the treatment electrode;
3
Turn the spindle clockwise until it stops.
spin
clockwise
WARNING:
Do not apply excessive force when tightening the spindle. Do not use any tools to unscrew or screw the spindle.
1.
counterclockwise
spin
2.
a new electrode
insert
3.
1
2