1.0  Introduction

 

Introduction

 

PLASMA IQ utilizes a treatment method called plasma fulguration, which causes controlled skin damage. It offers a low risk 
procedure, with minimal side effects. The device works through the generation of an electrical arc, which causes 
epidermal fulguration, with minimal effect on surrounding tissue. The PLASMA IQ works by creating an arc between the 
device tip and the skin, by ionizing the gas particles in the air. The resultant production of plasma causes fulguration of the 
epidermis of the skin; turning a solid straight into gas. It creates 1mm

or small treatment spots that show as tiny brown 

carbon crusts or scabs. These will remain for approximately a week before dropping off.

 

This Training Manual is only a means of support in the correct operation of the device. The PLASMA IQ device should 
only be used by medically licensed and certified practitioners who have undergone training conducted by authorized 
Neauvia trainers or distributors. The manufacturer shall not be liable for any potential damage resulting from improper 
operation of the device or treatments performed by untrained personnel who are not fully certified.

 

PLASMA IQ is used in the removal and destruction of skin lesions and the coagulation of tissue. 

PLASMA  IQ  is  Rx only.

 

WARNING:

 

PLASMA IQ DEVICE CAN CAUSE HARM TO THE USER AND/OR PATIENT IF NOT USED AS INSTRUCTED, USED FOR OTHER 
PURPOSES, OR CONTRADICTORY TO THE INTENDED METHOD OF OPERATION DEFINED IN THIS USER MANUAL. THE USER 
MUST FOLLOW TREATMENT PROCEDURES AND ABIDE BY THE GUIDELINES DEFINED IN THE USER MANUAL, AT ALL TIMES. 
THE PLASMA IQ DEVICE MAY ONLY BE USED WITH NEAUVIA PROVIDED ACTIVE ELECTRODES.

 

1

 

Summary of Contents for PLASMA IQ

Page 1: ...TRAINING MANUAL Neauvia North America Customer Service 1 866 836 3113...

Page 2: ...NTS 1 INTRODUCTION 3 2 PATIENT CONSULTATION 4 3 PREPARING THE DEVICE 6 4 PRE TREAMENT CARE 7 5 TECHNIQUE 8 5 1 Zig zag pattern technique 8 6 SKIN LESION 10 7 POST TREATMENT RECOMMENDATIONS 11 8 DEVICE...

Page 3: ...is only a means of support in the correct operation of the device The PLASMA IQ device should only be used by medically licensed and certified practitioners who have undergone training conducted by a...

Page 4: ...the orbital rim which includes the upper and lower eyelids Discuss contraindications with your patient and ensure that he she doesn t have any pre existing conditions that would make a treatment ineff...

Page 5: ...5 Swelling bruising Scabbing crusting 2 0 Patient Consultation Skin discoloration Scarring Redness tenderness RISKS OF PLASMA IQ TREATMENT...

Page 6: ...for What are the stages of thetreatments What types of sensations can they expect throughout the treatment What they should expect immediately after the treatment What kind of after care they should f...

Page 7: ...device there are two buttons used to provide two different output voltages marked with inscriptions H high and L low Figure 1 Device PLASMA IQ Preparing the device and equipment for work Before starti...

Page 8: ...8 4 0 Pre Treatment Care Pre Treatment Recommendations Avoid sun exposure or tanning beds for 2 4 weeks before treatment Ideally begin applying SPF 50 to prep skin for 2 4 weeks prior to treatment 1...

Page 9: ...as an initial measure before increasing voltage for more intensive fulguration It is recommended to use the LOW setting whenever possible for patient safety and comfort button LOW 650 V Controlled fu...

Page 10: ...skin using gauze and a non alcohol antiseptic 4 Dry area completely with gauze 5 Attach single use disposable electrode to the PLASMA IQ device as instructed on page 6 6 Press L button to set voltage...

Page 11: ...e effect 2 Carefully remove numbingcream 3 Cleanse the skin using gauze and a non alcohol antiseptic 4 Dry area completely with gauze 5 Attach single use disposable electrode to the PLASMA IQ device a...

Page 12: ...is normal and typically only lasts for about an hour though some variation in time may occur Continue to apply moisturizer such as petroleum jelly twice a day after gentle cleansing Additional moistu...

Page 13: ...t position as described in the manual If the electrode is mounted incorrectly it may affect the performance of the device WARNING Tighten the active electrode holder securely Improper tightening of th...

Page 14: ...visibly wet approximately 3 5 minutes Maintenance If you experience technical difficulty with your device contact Neauvia North America at 1 866 836 3113 or email service neauvia us com PLASMA IQ dev...

Page 15: ...14...

Page 16: ...15...

Page 17: ...Neauvia North America Customer Service 1 866 836 3113 2019 Neauvia North America 8480 Honeycutt Road Raleigh NC 27615 neauvia com PI NAM 01 190005 REV Code 122021...

Reviews: