background image

 

10

 

 

5 Operating 

Modes 

 

5.1 REBOUND 

 
 
 
 
 
 

 

 

 

 
 
 
 
 
 
 
When this mode is selected an operator may cock the hammer and take 
impact readings.  The display will indicate the rebound value and the number 
of impacts taken.  Up to 20 impacts can be taken for a given Test Record.  If 
the operator requires less than 20 impacts he can terminate the Test Record 
by pressing the Enter Key.  The display will momentarily display “Saving Test“ 
and then display the Average rebound value.  The operator may view the 
Median rebound value by pressing the Up Arrow Key.  Pressing the Down 
Arrow Key completes the Test Record and the display will be as indicated 
above.  The Test Record Number will be automatically incremented to the 
next number. 

T e s t  2 3 # I  7   h h :  m m 

 

 

 

 

 

 

 

 

Test Record Number 
0 - 936 

Number of Impacts 
0 - 20 

Current Time 

Rebound Value 

Current Date 

R: Rebound 

Summary of Contents for W-D-2005

Page 1: ...W D 2005 W D 1705 W D 1505 Digital Hardness tester Instruction Manual James Instruments Inc 3727 N Kedzie Avenue Chicago IL 60618 Phone 773 463 6565 1 800 426 6500 Fax 773 463 0009 www ndtjames com...

Page 2: ...RROW 9 4 4 ENTER KEY 9 4 5 ESC KEY 9 4 5 1 Exiting Print Mode 9 4 5 2 Entering Test Routines 9 4 5 3 Exiting Test Mode 9 5 OPERATING MODES 10 5 1 REBOUND 10 5 2 ANGLE 11 5 3 CONFIG 11 5 4 UPLOAD DATA...

Page 3: ...HE FACTORY AND IS LEFT HERE FOR REFERENCE PURPOSES ONLY 16 7 1 INITIALIZE MEMORY AND TEST RECORD NUMBER 16 7 2 SET HAMMER ZERO 16 7 3 CALIBRATING THE HAMMER 16 7 4 INSTRUMENT CONFIGURATION 16 8 MAINTE...

Page 4: ...ken into account the mentioned differences in capabilities between both systems this instruction manual covers the required material for the proper use of the W D 1505 system The hardness tester is an...

Page 5: ...Do not touch the side button whilst pressing on the piston On removing the device from the surface being tested the piston once more emerges completely from the device which is immediately ready for a...

Page 6: ...sting in floors or slabs type of structures or against any horizontal surface 0 degrees when testing against walls or vertical concrete members and finally 90 degrees when testing in an overhead posit...

Page 7: ...ted area Positioning the hammer perpendicular to the surface and deploy the impact plunger pushing against the surface until the push button spring out Caution the hammer should always be held in both...

Page 8: ...to reset the Test Number to zero UPLOAD DATA This mode is used to send stored Test Records to a PC Computer PRINT The Print mode is used to select and print a specific Test Record SET CLOCK This mode...

Page 9: ...pending on the mode that the instrument is in 4 5 1 Exiting Print Mode The ESC Key is used for two different operating conditions within the printing mode If the print mode is selected and no Test Rec...

Page 10: ...Test Record by pressing the Enter Key The display will momentarily display Saving Test and then display the Average rebound value The operator may view the Median rebound value by pressing the Up Arro...

Page 11: ...he Test Record Using the Up and Down Keys select the rebound discard value and press the Enter Key Displayed then will be Test Num Reset N This configuration item is used to reset the Test Record Numb...

Page 12: ...D TYPE Rebound UNITS Number DISCARD 15 IMPACT DATA Impact 1 42 Impact 2 40 Impact 3 45 Impact 4 50 Impact 5 41 Impact 6 40 Impact 7 40 Impact 8 42 COMPUTED DATA AVERAGE 43 MEDIAN 45 MINIMUM 40 MAXIMUM...

Page 13: ...t without printing 5 6 SET CLOCK This mode is used to set the internal Real Time Clock This device keeps track of the current time and date A battery is attached to this device so that even with power...

Page 14: ...and 3 together is connected to the upload serial port If this plug is not install and error message will be displayed A string of characters COMM Test 1234 are transmitted and then received The recei...

Page 15: ...TER Key Displayed will be Mem Test Reset 6 7 TEST ADC This mode is used to test the functionality of the potentiometer mounted on the hammerhead and the Analog to Digital circuitry When this mode is s...

Page 16: ...MER ZERO and press the ENTER Key Displayed will be Cock Hammer Head Press Enter Following these instructions cock and lock the hammer or hammer fully deployed and then press the ENTER Key Displayed wi...

Page 17: ...l as 68 8 2 CLEANING THE INSTRUMENT The impact plunger and housing should be wiped using a dry and clean cloth after using the instrument Avoid the use of abrasives or solvents to clean the hammer and...