background image

 

UVe-LUX Operators Manual ver 1_02.docx 

24/11/2009 

 

2.5

 

Display 

The meter is fitted with an OLED display that gives a much higher 
contrast and superior viewing angle than traditional TFT or LCD 
displays.  The display is protected behind a protective plastic cover, 
but care should still be taken to protect the screen area from blows or 
scratches or other damage. 

In normal operation when the unit is powered up, the first screen to 
appear is the Logo screen, followed by the Splash screen.  The Logo 

screen contains information about the meter, including the serial 
number and date of last calibration for the unit.  

Keeping the ON/OFF 

button held down during power on will pause the meter at the splash screen until the button is 
released.

 

 

 

 

  

2.5.1

 

Measurement Display 

When the unit is powered up it will normally enter the 
measurement display after the Splash Screen. 

The measurement display contains five components:- 

 

Battery level – This gives an indication of the state 
of the battery life. 

 

WL reading- this is the current reading  for the 
white light sensor (this reading is rounded). 

 

WL max – this is the maximum value the meter 
has observed for white light (this reading is 
absolute).  This reading is reset when the 
ON/OFF button is pressed. 

 

UV reading- this is the current reading for the UV sensor (this reading is rounded). 

 

UV max – this is the maximum value the meter has observed for UV (this reading is 
absolute).  This reading is reset when the ON/OFF button is pressed. 

If the meter is exposed to light levels greater than it’s designed range the meter will saturate

.  When 

the meter is saturated it will display “>5000” “>10000” or “++++” in red to indicate that it has 
saturated.

   The meter will return to normal operation once the light levels have returned to the 

meters design range. 

Figure 6 Viewing angle 

Figure 9 Measurement Display 

Figure 8 Splash Screen 

Figure 7 Logo Screen 

Summary of Contents for UVe-LUX

Page 1: ...Ver 1 02 NDT Consultants Ltd Middlemarch House Siskin Drive Coventry 44 0 2476 511151 sales ndt consultants co uk 24 11 2009 UVe LUX Operators Manual...

Page 2: ...4 11 2009 Revision History Version Date Page Description 1 00 20 11 09 All 1st draft 1 01 23 11 09 1 4 Typos and spelling mistakes corrected 1 02 24 11 09 1 5 References to visible light replaced with...

Page 3: ...l ver 1_02 docx iii 24 11 2009 Table of Contents 1 Introduction 1 2 System Components 1 2 1 Sensor housing 1 2 2 Meter housing 2 2 3 Buttons 2 2 4 The Battery 3 2 5 Display 4 2 5 1 Measurement Display...

Page 4: ...WL 0 1 to 5000 lux UVA 1 to 10 000 W cm Accuracy WL better than 5 UV better than 5 2 System Components The system consists of a sensor housing containing both the WL and UV sensors and the meter hous...

Page 5: ...eadings Pressing this button while the system is in menu mode will select the currently highlighted menu function Down button Pressing the down button in measurement mode makes the screen darker by on...

Page 6: ...orientation with the positive terminal of the battery being connected to the clip in the battery compartment The battery should only be fitted in the orientation described if the battery is forced in...

Page 7: ...nt display after the Splash Screen The measurement display contains five components Battery level This gives an indication of the state of the battery life WL reading this is the current reading for t...

Page 8: ...me screen has appeared Once in menu mode the UP and DOWN keys can used to scroll through menu items and pressing the ON OFF key is used to select menu items Holding the ON OFF key down for several sec...

Page 9: ...Settings menu Set auto off Sets the auto power off time between 20s and 6000s Set units Enters Set Units menu LUX Selects LUX as white light units Foot Candles Sets Foot Candles as white light units...

Reviews: